Racial effects on masimo pulse oximetry: impact of low perfusion index.
Sharma, Vikrant; Barker, Steven J; Sorci, Rebecca; Park, Linus; Wilson, William C.
Afiliación
  • Sharma V; Vice President, Optical Sciences, Masimo Corp, Irvine, CA, USA.
  • Barker SJ; Chief Science Officer, Masimo Corp, Irvine, CA, USA.
  • Sorci R; Director, Clinical Publication Review and Communications, Masimo Corp, Irvine, CA, USA.
  • Park L; Vice President Regulatory Affairs, Masimo Corp, Irvine, CA, USA.
  • Wilson WC; Executive Vice President, Clinical Research Operations and Medical Affairs, Masimo Corp, Irvine, CA, USA. william.wilson@masimo.com.
J Clin Monit Comput ; 38(2): 347-354, 2024 Apr.
Article en En | MEDLINE | ID: mdl-38238634
ABSTRACT

PURPOSE:

 Evaluate the SpO2-SaO2 difference between Black and White volunteer subjects having a low perfusion index (Pi) compared to those having a normal Pi.

METHODS:

 The Pi data were abstracted from electronic files collected on 7183 paired SpO2-SaO2 samples (3201 Black and 3982 White) from a recently reported desaturation study of 75 subjects (39 Black and 36 White) where SaO2 values were sequentially decreased from 100 to 70%. The Pi values from that dataset were divided into two groups (Pi ≤ 1 or Pi > 1) for analysis. A Pi value ≤ 1 was considered "low perfusion" and a Pi value > 1 was considered "normal perfusion". Statistical calculations included values of bias (mean difference of SpO2-SaO2), precision (standard deviation of the difference), and accuracy (root-mean-square error [ARMS]). During conditions of low perfusion (Pi ≤ 1, range [0.1 to 1]), overall bias and precision were + 0.48% ± 1.59%, while bias and precision were + 0.19 ± 1.53%, and + 0.91 ± 1.57%, for Black and White subjects, respectively.

RESULTS:

 During normal perfusion (Pi > 1, range [1 to 12]), overall bias and precision were + 0.18% ± 1.34%, while bias and precision were -0.26 ± 1.37%, and - 0.12 ± 1.31%, for Black and White subjects, respectively. ARMS was 1.37% in all subjects with normal perfusion and 1.64% in all subjects with low perfusion.

CONCLUSION:

 Masimo SET® pulse oximeters with RD SET® sensors are accurate for individuals of both Black and White races when Pi is normal, as well as during conditions when Pi is low. The ARMS for all conditions studied is well within FDA standards. This study was conducted in healthy volunteers during well-controlled laboratory desaturations, and results could vary under certain challenging clinical conditions.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Oximetría / Índice de Perfusión Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: J Clin Monit Comput Asunto de la revista: INFORMATICA MEDICA / MEDICINA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Oximetría / Índice de Perfusión Tipo de estudio: Guideline Límite: Humans Idioma: En Revista: J Clin Monit Comput Asunto de la revista: INFORMATICA MEDICA / MEDICINA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos
...