Optimizing High-Resolution MR Angiography: The Synergistic Effects of 3D Wheel Sampling and Deep Learning-Based Reconstruction.
Sasaki, Goh; Uetani, Hiroyuki; Nakaura, Takeshi; Nakahara, Keiichi; Morita, Kosuke; Nagayama, Yasunori; Kidoh, Masafumi; Iwashita, Koya; Yoshida, Naofumi; Hokamura, Masamichi; Yamashita, Yuichi; Nakajima, Makoto; Ueda, Mitsuharu; Hirai, Toshinori.
Afiliación
  • Sasaki G; From the Departments of Diagnostic Radiology, and.
  • Uetani H; From the Departments of Diagnostic Radiology, and.
  • Nakaura T; From the Departments of Diagnostic Radiology, and.
  • Nakahara K; Neurology, Graduate School of Medical Sciences, Kumamoto University.
  • Morita K; Central Radiology Section, Kumamoto University Hospital, Kumamoto.
  • Nagayama Y; From the Departments of Diagnostic Radiology, and.
  • Kidoh M; From the Departments of Diagnostic Radiology, and.
  • Iwashita K; From the Departments of Diagnostic Radiology, and.
  • Yoshida N; From the Departments of Diagnostic Radiology, and.
  • Hokamura M; From the Departments of Diagnostic Radiology, and.
  • Yamashita Y; MRI Clinical Strategy Group, MRI Sales Department, Canon Medical Systems Corporation, Kanagawa, Japan.
  • Nakajima M; Neurology, Graduate School of Medical Sciences, Kumamoto University.
  • Ueda M; Neurology, Graduate School of Medical Sciences, Kumamoto University.
  • Hirai T; From the Departments of Diagnostic Radiology, and.
Article en En | MEDLINE | ID: mdl-38346820
ABSTRACT

OBJECTIVE:

The aim of this study was to assess the utility of the combined use of 3D wheel sampling and deep learning-based reconstruction (DLR) for intracranial high-resolution (HR)-time-of-flight (TOF)-magnetic resonance angiography (MRA) at 3 T.

METHODS:

This prospective study enrolled 20 patients who underwent head MRI at 3 T, including TOF-MRA. We used 3D wheel sampling called "fast 3D" and DLR for HR-TOF-MRA (spatial resolution, 0.39 × 0.59 × 0.5 mm3) in addition to conventional MRA (spatial resolution, 0.39 × 0.89 × 1 mm3). We compared contrast and contrast-to-noise ratio between the blood vessels (basilar artery and anterior cerebral artery) and brain parenchyma, full width at half maximum in the P3 segment of the posterior cerebral artery among 3 protocols. Two board-certified radiologists evaluated noise, contrast, sharpness, artifact, and overall image quality of 3 protocols.

RESULTS:

The contrast and contrast-to-noise ratio of fast 3D-HR-MRA with DLR are comparable or higher than those of conventional MRA and fast 3D-HR-MRA without DLR. The full width at half maximum was significantly lower in fast 3D-MRA with and without DLR than in conventional MRA (P = 0.006, P < 0.001). In qualitative evaluation, fast 3D-MRA with DLR had significantly higher sharpness and overall image quality than conventional MRA and fast 3D-MRA without DLR (sharpness P = 0.021, P = 0.001; overall image quality P = 0.029, P < 0.001).

CONCLUSIONS:

The combination of 3D wheel sampling and DLR can improve visualization of arteries in intracranial TOF-MRA.

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Observational_studies / Qualitative_research Idioma: En Revista: J Comput Assist Tomogr Año: 2024 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Tipo de estudio: Observational_studies / Qualitative_research Idioma: En Revista: J Comput Assist Tomogr Año: 2024 Tipo del documento: Article
...