Framework to optimize fixed-length micro-CT systems for propagation-based phase-contrast imaging.
Opt Express
; 32(4): 4839-4856, 2024 Feb 12.
Article
en En
| MEDLINE
| ID: mdl-38439226
ABSTRACT
A laboratory X-ray imaging system with a setup that closely resembles commercial micro-CT systems with a fixed source-to-detector distance of â¼90â
cm is investigated for single distance propagation-based phase-contrast imaging and computed tomography (CT). The system had a constant source-to-detector distance, and the sample positions were optimized. Initially, a PTFE wire was imaged, both in 2D and 3D, to characterize fringe contrast and spatial resolution for different X-ray source settings and source-to-sample distances. The results were compared to calculated values based on theoretical models and to simulated (wave-optics based) results, with good agreement being found. The optimization of the imaging system is discussed. CT scans of two biological samples, a tissue-engineered esophageal scaffold and a rat heart, were then acquired at the optimum parameters, demonstrating that significant image quality improvements can be obtained with widely available components placed inside fixed-length cabinets through proper optimization of propagation-based phase-contrast.