Your browser doesn't support javascript.
loading
A novel intracoronary hypothermia device reduces myocardial reperfusion injury in pigs.
Pei, Zhiqiang; Qiu, Jin; Zhao, Yongchao; Song, Shuai; Wang, Rui; Luo, Wei; Cai, Xingxing; Liu, Bin; Chen, Han; Yin, Jiasheng; Weng, Xinyu; Wu, Yizhe; Li, Chenguang; Shen, Li; Ge, Junbo.
Afiliación
  • Pei Z; Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai 200032, China.
  • Qiu J; Shanghai Institute of Cardiovascular Diseases, Shanghai 200032, China.
  • Zhao Y; Department of Cardiology, Taiyuan Central Hospital of Shanxi Medical University, Taiyuan, Shanxi 030009, China.
  • Song S; Department of Cardiology, Affiliated Hospital of Zunyi Medical University, Zunyi, Guizhou 563000, China.
  • Wang R; Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai 200032, China.
  • Luo W; Shanghai Institute of Cardiovascular Diseases, Shanghai 200032, China.
  • Cai X; National Clinical Research for Interventional Medicine, Shanghai 200032, China.
  • Liu B; Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai 200032, China.
  • Chen H; Shanghai Institute of Cardiovascular Diseases, Shanghai 200032, China.
  • Yin J; National Clinical Research for Interventional Medicine, Shanghai 200032, China.
  • Weng X; Department of Cardiology, Zhongshan Hospital, Fudan University, Shanghai 200032, China.
  • Wu Y; Shanghai Institute of Cardiovascular Diseases, Shanghai 200032, China.
  • Li C; National Clinical Research for Interventional Medicine, Shanghai 200032, China.
  • Shen L; Department of Cardiology, Xinhua Hospital Affiliated to Shanghai Jiaotong University School of Medicine, Shanghai 201322, China.
  • Ge J; Department of Cardiology, Taiyuan Central Hospital of Shanxi Medical University, Taiyuan, Shanxi 030009, China.
Chin Med J (Engl) ; 137(20): 2461-2472, 2024 Oct 20.
Article en En | MEDLINE | ID: mdl-38445387
ABSTRACT

BACKGROUND:

Hypothermia therapy has been suggested to attenuate myocardial necrosis; however, the clinical implementation as a valid therapeutic strategy has failed, and new approaches are needed to translate into clinical applications. This study aimed to assess the feasibility, safety, and efficacy of a novel selective intracoronary hypothermia (SICH) device in mitigating myocardial reperfusion injury.

METHODS:

This study comprised two phases. The first phase of the SICH was performed in a normal porcine model for 30 minutes ( n = 5) to evaluate its feasibility. The second phase was conducted in a porcine myocardial infarction (MI) model of myocardial ischemia/reperfusion which was performed by balloon occlusion of the left anterior descending coronary artery for 60 minutes and maintained for 42 days. Pigs in the hypothermia group ( n = 8) received hypothermia intervention onset reperfusion for 30 minutes and controls ( n = 8) received no intervention. All animals were followed for 42 days. Cardiac magnetic resonance analysis (five and 42 days post-MI) and a series of biomarkers/histological studies were performed.

RESULTS:

The average time to lower temperatures to a steady state was 4.8 ± 0.8 s. SICH had no impact on blood pressure or heart rate and was safely performed without complications by using a 3.9 F catheter. Interleukin-6 (IL-6), tumor necrosis factor-α, C-reactive protein (CRP), and brain natriuretic peptide (BNP) were lower at 60 min post perfusion in pigs that underwent SICH as compared with the control group. On day 5 post MI/R, edema, intramyocardial hemorrhage, and microvascular obstruction were reduced in the hypothermia group. On day 42 post MI/R, the infarct size, IL-6, CRP, BNP, and matrix metalloproteinase-9 were reduced, and the ejection fraction was improved in pigs that underwent SICH.

CONCLUSIONS:

The SICH device safely and effectively reduced the infarct size and improved heart function in a pig model of MI/R. These beneficial effects indicate the clinical potential of SICH for treatment of myocardial reperfusion injury.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Daño por Reperfusión Miocárdica / Hipotermia Inducida Límite: Animals Idioma: En Revista: Chin Med J (Engl) Año: 2024 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Daño por Reperfusión Miocárdica / Hipotermia Inducida Límite: Animals Idioma: En Revista: Chin Med J (Engl) Año: 2024 Tipo del documento: Article País de afiliación: China
...