Validation of 2D flow MRI for helical and vortical flows.
Mehmood, Zia; Assadi, Hosamadin; Grafton-Clarke, Ciaran; Li, Rui; Matthews, Gareth; Alabed, Samer; Girling, Rebekah; Underwood, Victoria; Kasmai, Bahman; Zhao, Xiaodan; Ricci, Fabrizio; Zhong, Liang; Aung, Nay; Petersen, Steffen Erhard; Swift, Andrew J; Vassiliou, Vassilios S; Cavalcante, João; Geest, Rob J van der; Garg, Pankaj.
Afiliación
  • Mehmood Z; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Assadi H; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Grafton-Clarke C; Department of Cardiovascular and Metabolic Health, University of East Anglia Norwich Medical School, Norwich, UK.
  • Li R; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Matthews G; Department of Cardiovascular and Metabolic Health, University of East Anglia, Norwich, UK.
  • Alabed S; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Girling R; Department of Cardiovascular and Metabolic Health, University of East Anglia, Norwich, UK.
  • Underwood V; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Kasmai B; Department of Cardiovascular and Metabolic Health, University of East Anglia, Norwich, UK.
  • Zhao X; Department of Infection, Immunity & Cardiovascular Disease, University of Sheffield, Sheffield, UK.
  • Ricci F; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Zhong L; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Aung N; Norfolk and Norwich University Hospitals NHS Foundation Trust, Norwich, UK.
  • Petersen SE; Department of Cardiovascular and Metabolic Health, University of East Anglia, Norwich, UK.
  • Swift AJ; National Heart Centre Singapore, Singapore.
  • Vassiliou VS; Neuroscience, Imaging and Clinical Sciences, Gabriele d'Annunzio University of Chieti and Pescara, Chieti Scalo, Italy.
  • Cavalcante J; National Heart Centre Singapore, Singapore.
  • Geest RJV; Queen Mary University of London, London, UK.
  • Garg P; Advanced Cardiovascular Imaging William Harvey Research Institute, The London Chest Hospital, London, UK.
Open Heart ; 11(1)2024 Mar 08.
Article en En | MEDLINE | ID: mdl-38458769
ABSTRACT

PURPOSE:

The main objective of this study was to develop two-dimensional (2D) phase contrast (PC) methods to quantify the helicity and vorticity of blood flow in the aortic root.

METHODS:

This proof-of-concept study used four-dimensional (4D) flow cardiovascular MR (4D flow CMR) data of five healthy controls, five patients with heart failure with preserved ejection fraction and five patients with aortic stenosis (AS). A PC through-plane generated by 4D flow data was treated as a 2D PC plane and compared with the original 4D flow. Visual assessment of flow vectors was used to assess helicity and vorticity. We quantified flow displacement (FD), systolic flow reversal ratio (sFRR) and rotational angle (RA) using 2D PC.

RESULTS:

For visual vortex flow presence near the inner curvature of the ascending aortic root on 4D flow CMR, sFRR demonstrated an area under the curve (AUC) of 0.955, p<0.001. A threshold of >8% for sFRR had a sensitivity of 82% and specificity of 100% for visual vortex presence. In addition, the average late systolic FD, a marker of flow eccentricity, also demonstrated an AUC of 0.909, p<0.001 for visual vortex flow. Manual systolic rotational flow angle change (ΔsRA) demonstrated excellent association with semiautomated ΔsRA (r=0.99, 95% CI 0.9907 to 0.999, p<0.001). In reproducibility testing, average systolic FD (FDsavg) showed a minimal bias at 1.28% with a high intraclass correlation coefficient (ICC=0.92). Similarly, sFRR had a minimal bias of 1.14% with an ICC of 0.96. ΔsRA demonstrated an acceptable bias of 5.72°-and an ICC of 0.99.

CONCLUSION:

2D PC flow imaging can possibly quantify blood flow helicity (ΔRA) and vorticity (FRR). These imaging biomarkers of flow helicity and vorticity demonstrate high reproducibility for clinical adoption. TRIALS REGISTRATION NUMBER NCT05114785.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Imagen por Resonancia Magnética Límite: Humans Idioma: En Revista: Open Heart Año: 2024 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Imagen por Resonancia Magnética Límite: Humans Idioma: En Revista: Open Heart Año: 2024 Tipo del documento: Article
...