A wearable flexible electrochemical biosensor with CuNi-MOF@rGO modification for simultaneous detection of uric acid and dopamine in sweat.
Anal Chim Acta
; 1299: 342441, 2024 Apr 22.
Article
en En
| MEDLINE
| ID: mdl-38499429
ABSTRACT
BACKGROUND:
In health assessment and personalized medical services, accurate detection of biological markers such as dopamine (DA) and uric acid (UA) in sweat is crucial for providing valuable physiological information. However, there are challenges in detecting sweat biomarkers due to their low concentrations, variations in sweat yield among individuals, and the need for efficient sweat collection.RESULTS:
We synthesized CuNi-MOF@rGO as a high-activity electrocatalyst and investigated its feasibility and electrochemical mechanism for simultaneously detecting low-concentration biomarkers UA and DA. Interaction between the non-coordinating carboxylate group and the sample produces effective separation signals for DA and UA. The wearable biomimetic biosensor has a wide linear range of 1-500 µM, with a detection limit of 9.41 µM and sensitivity of 0.019 µA µM-1 cm-2 for DA, and 10-1000 µM, with a detection limit of 9.09 µM and sensitivity of 0.026 µA µM-1 cm-2 for UA. Thus, our sensor performs excellently in detecting low-concentration biomarkers. To improve sweat collection, we designed a microfluidic-controlled device with hydrophilic modification in the microchannel. Experimental results show optimal ink flow at 2% concentration. Overall, we developed an innovative and highly active electrocatalyst, successfully enabling simultaneous detection of low-concentration biomarkers UA and DA.SIGNIFICANCE:
This study provides a strategy for sweat analysis and health monitoring. Moreover, the sensor also showed good performance in detecting real sweat samples. This study has shown great potential in future advances in sweat analysis and health monitoring.Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Técnicas Biosensibles
/
Dispositivos Electrónicos Vestibles
Límite:
Humans
Idioma:
En
Revista:
Anal Chim Acta
Año:
2024
Tipo del documento:
Article