Cardiac implantable electronic device-induced tricuspid regurgitation: Implications and management.
Murphy, Sean P; Lew, Jeanney; Yucel, Evin; Singh, Jagmeet; Mela, Theofanie.
Afiliación
  • Murphy SP; Division of Cardiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Lew J; Cardiac Ultrasound Laboratory, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Yucel E; Division of Cardiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Singh J; The Demoulas Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, Massachusetts, USA.
  • Mela T; Division of Cardiology, Massachusetts General Hospital, Boston, Massachusetts, USA.
J Cardiovasc Electrophysiol ; 35(5): 1017-1025, 2024 May.
Article en En | MEDLINE | ID: mdl-38501386
ABSTRACT
Tricuspid regurgitation (TR) secondary to cardiac implantable electronic devices (CIEDs) has been well documented and is associated with worse cardiovascular outcomes. A variety of mechanisms have been proposed including lead-induced mechanical disruption of the tricuspid valvular or subvalvular apparatus and pacing-induced electrical dyssynchrony. Patient characteristics such as age, sex, baseline atrial fibrillation, and pre-existing TR have not been consistent predictors of CIED-induced TR. While two-dimensional echocardiography is helpful in assessing the severity of TR, three-dimensional echocardiography has significantly improved accuracy in identifying the etiology of TR and whether lead position contributes to TR. Three-dimensional echocardiography may therefore play a future role in optimizing lead positioning during implant to reduce the risk of CIED-induced TR. Optimal lead management strategies in addition to percutaneous interventions and surgery in alleviating TR are very important.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Insuficiencia de la Válvula Tricúspide / Desfibriladores Implantables Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Marcapaso Artificial / Insuficiencia de la Válvula Tricúspide / Desfibriladores Implantables Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos
...