Comparison of two self-expanding transcatheter heart valves for degenerated surgical bioprostheses: the AVENGER multicentre registry.
Kim, Won-Keun; Seiffert, Moritz; Rück, Andreas; Leistner, David M; Dreger, Henryk; Wienemann, Hendrik; Adam, Matti; Möllmann, Helge; Blumenstein, Johannes; Eckel, Clemens; Buono, Andrea; Maffeo, Diego; Messina, Antonio; Holzamer, Andreas; Sossalla, Samuel; Costa, Giuliano; Barbanti, Marco; Motta, Silvia; Tamburino, Corrado; von der Heide, Ina; Glasmacher, Julius; Sherif, Mohammad; Seppelt, Philipp; Fichtlscherer, Stephan; Walther, Thomas; Castriota, Fausto; Nerla, Roberto; Frerker, Christian; Schmidt, Tobias; Wolf, Alexander; Adamaszek, Martin M; Giannini, Francesco; Vanhaverbeke, Maarten; Van de Walle, Stefaan; Stammen, Francis; Toggweiler, Stefan; Brunner, Stephanie; Mangieri, Antonio; Gitto, Mauro; Kaleschke, Gerrit; Ninios, Vlasis; Ninios, Ilias; Hübner, Judith; Xhepa, Erion; Renker, Matthias; Charitos, Efstratios I; Joner, Michael; Rheude, Tobias.
Afiliación
  • Kim WK; Department of Cardiology, Kerckhoff Heart Center, Bad Nauheim, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Rhein/Main, Germany.
  • Seiffert M; Department of Cardiac Surgery, Kerckhoff Heart Center, Bad Nauheim, Germany.
  • Rück A; Department of Cardiology, Justus-Liebig University of Gießen, Gießen, Germany.
  • Leistner DM; Department of Cardiology, University Heart and Vascular Center Hamburg, University Hospital Hamburg-Eppendorf (UKE), Hamburg, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Hamburg/Kiel/Lübeck, Germany.
  • Dreger H; Department of Cardiology, Karolinska University Hospital, Stockholm, Sweden.
  • Wienemann H; University Heart & Vascular Center Frankfurt, Frankfurt, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Rhein/Main, Germany.
  • Adam M; Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum der Charite (DHZC), Berlin, Germany.
  • Möllmann H; Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum der Charite (DHZC), Berlin, Germany.
  • Blumenstein J; Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
  • Eckel C; Clinic III for Internal Medicine, Faculty of Medicine and University Hospital Cologne, Cologne, Germany.
  • Buono A; Clinic III for Internal Medicine, Faculty of Medicine and University Hospital Cologne, Cologne, Germany.
  • Maffeo D; Department of Cardiology, St. Johannes-Hospital, Dortmund, Germany.
  • Messina A; Department of Cardiology, St. Johannes-Hospital, Dortmund, Germany.
  • Holzamer A; Department of Cardiology, Carl-von-Ossietzky University Oldenburg, Oldenburg, Germany.
  • Sossalla S; Department of Cardiology, St. Johannes-Hospital, Dortmund, Germany.
  • Costa G; Department of Cardiology, Carl-von-Ossietzky University Oldenburg, Oldenburg, Germany.
  • Barbanti M; Cardiovascular Department, Interventional Cardiology Unit, Fondazione Poliambulanza Istituto Ospedaliero, Brescia, Italy.
  • Motta S; Cardiovascular Department, Interventional Cardiology Unit, Fondazione Poliambulanza Istituto Ospedaliero, Brescia, Italy.
  • Tamburino C; Department of Cardiothoracic Surgery, Fondazione Poliambulanza Istituto Ospedaliero, Brescia, Italy and Operative Unit of Cardiac Surgery, Poliambulanza Foundation Ospital, Brescia, Italy.
  • von der Heide I; University Hospital of Regensburg, Medical Center, Regensburg, Germany.
  • Glasmacher J; Department of Cardiology, Kerckhoff Heart Center, Bad Nauheim, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Rhein/Main, Germany.
  • Sherif M; Department of Cardiology, Justus-Liebig University of Gießen, Gießen, Germany.
  • Seppelt P; Division of Cardiology, AOU Policlinico G. Rodolico-San Marco, Catania, Italy.
  • Fichtlscherer S; Università degli Studi Di Enna "Kore", Enna, Italy.
  • Walther T; Division of Cardiology, AOU Policlinico G. Rodolico-San Marco, Catania, Italy.
  • Castriota F; Division of Cardiology, AOU Policlinico G. Rodolico-San Marco, Catania, Italy.
  • Nerla R; Department of Cardiology, University Heart and Vascular Center Hamburg, University Hospital Hamburg-Eppendorf (UKE), Hamburg, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Hamburg/Kiel/Lübeck, Germany.
  • Frerker C; Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum der Charite (DHZC), Berlin, Germany.
  • Schmidt T; Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
  • Wolf A; Department of Cardiology, Angiology and Intensive Care Medicine, Deutsches Herzzentrum der Charite (DHZC), Berlin, Germany.
  • Adamaszek MM; Charité - Universitätsmedizin Berlin, corporate member of Freie Universität Berlin and Humboldt-Universität zu Berlin, Berlin, Germany.
  • Giannini F; University Heart & Vascular Center Frankfurt, Frankfurt, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Rhein/Main, Germany.
  • Vanhaverbeke M; University Heart & Vascular Center Frankfurt, Frankfurt, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Rhein/Main, Germany.
  • Van de Walle S; University Heart & Vascular Center Frankfurt, Frankfurt, Germany and DZHK (German Centre for Cardiovascular Research), Partner Site Rhein/Main, Germany.
  • Stammen F; Maria Cecilia Hospital, GVM Care & Research, Cotignola, Italy.
  • Toggweiler S; Maria Cecilia Hospital, GVM Care & Research, Cotignola, Italy.
  • Brunner S; Department of Cardiology, Universitätsklinikum Schleswig-Holstein, Campus Lübeck, Lübeck, Germany and DZHK (German Center for Cardiovascular Research), Partner Site Hamburg/Kiel/Lübeck, Germany.
  • Mangieri A; Department of Cardiology, Universitätsklinikum Schleswig-Holstein, Campus Lübeck, Lübeck, Germany and DZHK (German Center for Cardiovascular Research), Partner Site Hamburg/Kiel/Lübeck, Germany.
  • Gitto M; Contilia Herz- und Gefäßzentrum, Elisabeth-Krankenhaus Essen, Essen, Germany.
  • Kaleschke G; Contilia Herz- und Gefäßzentrum, Elisabeth-Krankenhaus Essen, Essen, Germany.
  • Ninios V; IRCCS Galeazzi Sant'Ambrogio Hospital, Milan, Italy.
  • Ninios I; Department of Cardiology, AZ Delta, Roeselare, Belgium.
  • Hübner J; Department of Cardiology, AZ Delta, Roeselare, Belgium.
  • Xhepa E; Department of Cardiology, AZ Delta, Roeselare, Belgium.
  • Renker M; Heart Center Lucerne, Lucerner Kantonsspital, Lucerne, Switzerland.
  • Charitos EI; Heart Center Lucerne, Lucerner Kantonsspital, Lucerne, Switzerland.
  • Joner M; Cardiocenter, IRCCS, Humanitas Research Hospital, Rozzano, Milan, Italy.
  • Rheude T; Department of Biomedical Sciences, Humanitas University, Pieve Emanuele, Milan, Italy and IRCCS Humanitas Research Hospital, Rozzano-Milan, Italy.
EuroIntervention ; 20(6): e363-e375, 2024 03 18.
Article en En | MEDLINE | ID: mdl-38506737
ABSTRACT

BACKGROUND:

There is a lack of comparative data on transcatheter aortic valve implantation (TAVI) in degenerated surgical prostheses (valve-in-valve [ViV]).

AIMS:

We sought to compare outcomes of using two self-expanding transcatheter heart valve (THV) systems for ViV.

METHODS:

In this retrospective multicentre registry, we included consecutive patients undergoing transfemoral ViV using either the ACURATE neo/neo2 (ACURATE group) or the Evolut R/PRO/PRO+ (EVOLUT group). The primary outcome measure was technical success according to Valve Academic Research Consortium (VARC)-3. Secondary outcomes were 30-day all-cause mortality, device success (VARC-3), coronary obstruction (CO) requiring intervention, rates of severe prosthesis-patient mismatch (PPM), and aortic regurgitation (AR) ≥moderate. Comparisons were made after 11 propensity score matching.

RESULTS:

The study cohort comprised 835 patients from 20 centres (ACURATE n=251; EVOLUT n=584). In the matched cohort (n=468), technical success (ACURATE 92.7% vs EVOLUT 88.9%; p=0.20) and device success (69.7% vs 73.9%; p=0.36) as well as 30-day mortality (2.8% vs 1.6%; p=0.392) were similar between the two groups. The mean gradients and rates of severe PPM, AR ≥moderate, or CO did not differ between the groups. Technical and device success were higher for the ACURATE platform among patients with a true inner diameter (ID) >19 mm, whereas a true ID ≤19 mm was associated with higher device success - but not technical success - among Evolut recipients.

CONCLUSIONS:

ViV TAVI using either ACURATE or Evolut THVs showed similar procedural outcomes. However, a true ID >19 mm was associated with higher device success among ACURATE recipients, whereas in patients with a true ID ≤19 mm, device success was higher when using Evolut.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Insuficiencia de la Válvula Aórtica / Bioprótesis / Oclusión Coronaria / Reemplazo de la Válvula Aórtica Transcatéter Límite: Humans Idioma: En Revista: EuroIntervention Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Año: 2024 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Insuficiencia de la Válvula Aórtica / Bioprótesis / Oclusión Coronaria / Reemplazo de la Válvula Aórtica Transcatéter Límite: Humans Idioma: En Revista: EuroIntervention Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / TERAPEUTICA Año: 2024 Tipo del documento: Article País de afiliación: Alemania
...