Whole Process of Standardization of Diffusion-Weighted Imaging: Phantom Validation and Clinical Application According to the QIBA Profile.
Choi, Se Jin; Kim, Kyung Won; Ko, Yousun; Cho, Young Chul; Jang, Ji Sung; Ahn, Hyemin; Kim, Dong Wook; Kim, Mi Young.
Afiliación
  • Choi SJ; Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Kim KW; Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Ko Y; Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Cho YC; Biomedical Research Center, Asan Institute for Life Sciences, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Jang JS; Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Ahn H; Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Kim DW; Department of Radiology and Research Institute of Radiology, Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
  • Kim MY; Department of Radiology, Asan Medical Institute of Convergence Science and Technology (AMIST), Asan Medical Center, University of Ulsan College of Medicine, Seoul 05505, Republic of Korea.
Diagnostics (Basel) ; 14(6)2024 Mar 09.
Article en En | MEDLINE | ID: mdl-38535004
ABSTRACT

Background:

To use the apparent diffusion coefficient (ADC) as reliable biomarkers, validation of MRI equipment performance and clinical acquisition protocols should be performed prior to application in patients. This study aims to validate various MRI equipment and clinical brain protocols for diffusion weighted imaging (DWI) using commercial phantom, and confirm the validated protocols in patients' images.

Methods:

The performance of four different scanners and clinical brain protocols were validated using a Quantitative Imaging Biomarker Alliance (QIBA) diffusion phantom and cloud-based analysis tool. We evaluated the performance metrics regarding accuracy and repeatability of ADC measurement using QIBA profile. The validated clinical brain protocols were applied to 17 patients, and image quality and repeatability of ADC were assessed.

Results:

The MRI equipment performance of all four MRI scanners demonstrated high accuracy in ADC measurement (ADC bias, -2.3% to -0.4%), excellent linear correlation to the reference ADC value (slope, 0.9 to 1.0; R2, 0.999-1.000), and high short-term repeatability [within-subject-coefficient-of-variation (wCV), 0% to 0.3%]. The clinical protocols were also validated by fulfilling QIBA claims with high accuracy (ADC bias, -3.1% to -0.7%) and robust repeatability (wCV, 0% to 0.1%). Brain DWI acquired using the validated clinical protocols showed ideal image quality (mean score ≥ 2.9) and good repeatability (wCV, 1.8-2.2).

Conclusions:

The whole process of standardization of DWI demonstrated the robustness of ADC with high accuracy and repeatability across diverse MRI equipment and clinical protocols in accordance with the QIBA claims.