Comparison of revision rates and epidemiological data of a single total knee arthroplasty system of different designs (cruciate retaining, posterior stabilized, mobile bearing, and fixed bearing): a meta-analysis and systematic review of clinical trials and national arthroplasty registries.
Arch Orthop Trauma Surg
; 144(5): 1997-2006, 2024 May.
Article
en En
| MEDLINE
| ID: mdl-38570357
ABSTRACT
BACKGROUND:
This study aimed to meta-analyze epidemiological data, revision rates, and incidences of different designs of a single Total Knee Arthroplasty System and compare these factors across different countries.METHODS:
A systematic review was conducted on clinical studies and arthroplasty registries of ATTUNE TKA from 1999 to 2020. The main endpoints analyzed were revision rates and epidemiological data.RESULTS:
The average age of patients was 67.8 years, with a gender distribution of 60% female and 40% male. The pooled average BMI was 29.4 kg/m2. Eight clinical studies showed a pooled revision rate per 100 observed CY of 0.5 (n = 1343 cases). Cumulative revision rates after 1, 3, and 5 years varied among registries, with the Swiss registry having the highest revision data (after 5 years 6.3%) and the American registry having the lowest revision data (after 5 years 1.7%). A comparison of the revision rates of mobile bearing and fixed bearing (41,200 cases) as well as cruciate retaining and posterior stabilized (n = 123,361 cases) showed no significant advantage in the first 5 years after implantation.CONCLUSION:
In conclusion, pooled data from 41,200 cases of TKA with a single Total Knee Arthroplasty System in two arthroplasty registries revealed that there was no significant difference in revision rates between the mobile bearing and fixed bearing design within the first 5 years after implantation. In addition, a comparison of the revision rates in n = 123,361 cases showed no significant advantage for cruciate retaining or posterior stabilized in the first 5 years after implantation.Palabras clave
Texto completo:
1
Colección:
01-internacional
Base de datos:
MEDLINE
Asunto principal:
Diseño de Prótesis
/
Reoperación
/
Sistema de Registros
/
Artroplastia de Reemplazo de Rodilla
/
Prótesis de la Rodilla
Límite:
Aged
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
Arch Orthop Trauma Surg
Año:
2024
Tipo del documento:
Article
País de afiliación: