Your browser doesn't support javascript.
loading
A novel XNA-based Luminex assay to detect UBA1 somatic mutations associated with VEXAS syndrome.
Ma, Yunqing; Hu, ShianPin; Ni, Rui; Liu, Wei; Fu, Andrew; Sha, Michael; Zhang, Aiguo; Lu, Chuanyi M.
Afiliación
  • Ma Y; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Hu S; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Ni R; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Liu W; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Fu A; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Sha M; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Zhang A; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
  • Lu CM; DiaCarta Inc., 4385 Hopyard Rd, Suite 100 Pleasanton, CA, 94588, USA.
Pract Lab Med ; 39: e00380, 2024 Mar.
Article en En | MEDLINE | ID: mdl-38715663
ABSTRACT

Objectives:

Patients with VEXAS syndrome carry mutations of UBA1 gene coding for the E1 enzyme. The three most frequent mutations are p.M41T(122T > C), p.M41V (c.121A > G), and p.M41L (c.121A > C) in codon 41 of exon 3. Currently, sanger sequencing was mainly used to detect these mutations, which has low sensitivity and low throughput. There is a need of high sensitivity, simple and high throughput method to characterize patients with VEXAS syndrome.

Methods:

Based on our proprietary XNA technology, we have developed a QClamp® Plex platform to detect eight mutations in a single reaction using the Luminex xMap technology. The assay sensitivity, specificity and precision were subsequently evaluated. Furthermore, the reference interval and clinical sensitivity/specificity were estimated using clinical healthy/positive DNA samples and the sanger sequencing method was used for comparison.

Results:

With spiking synthetic mutant DNA in wildtype GM24385 cell line DNA, this assay can detect UBA1 mutations with a detection limit of variant allele frequency (VAF) at 0.1-5%. Our assay shows 100% concordance with Sanger sequencing results when used for analyzing 15 positive and 19 negative clinical samples.

Conclusions:

The QClamps® Plex UBA1 Mutation Detection Assay is a quicker, simpler, and more sensitive assay that can accurately detect the UBA1 mutations even at early stages with low mutation frequency.