Your browser doesn't support javascript.
loading
May the 4th be with you: mixed-methods best-evidence synthesis on 4th-generation alumina-zirconia ceramic bearings in total hip arthroplasty.
Perka, Carsten; Mohaddes, Maziar; Zagra, Luigi; Ekkernkamp, Axel; Keller, Niklas; Stengel, Dirk.
Afiliación
  • Perka C; Center for Musculoskeletal Surgery, Charité Medical University Center, Berlin, Germany.
  • Mohaddes M; Hässleholms Hospital, Region Skåne, Hässleholm, Sweden.
  • Zagra L; Orthopedics, Faculty of Medicine, Department of Clinical Sciences, Lund University, Malmö, Sweden.
  • Ekkernkamp A; Hip Department, IRCCS Istituto Ortopedico Galeazzi, Milan, Italy.
  • Keller N; BG Klinikum Unfallkrankenhaus Berlin gGmbH, Berlin, Germany.
  • Stengel D; BG Kliniken - Klinikverbund der gesetzlichen Unfallversicherung gGmbH, Berlin, Germany.
EFORT Open Rev ; 9(7): 632-645, 2024 Jul 01.
Article en En | MEDLINE | ID: mdl-38949170
ABSTRACT

Purpose:

To assess utility, benefits, and risks of 4th-generation alumina-zirconia ceramic pairings in elective total hip arthroplasty (THA).

Methods:

A comprehensive mixed-methods best-evidence synthesis using data from systematic reviews, randomized controlled trials (RCTs), prospective and retrospective cohort studies, as well as joint replacement registries, was conducted to estimate overall revision and survival rates, periprosthetic infection, bearing fractures, and noise phenomena with 4th-generation alumina-zirconia ceramic versus other tribological couplings in elective THA. The systematic review part across multiple databases was registered with PROSPERO (CRD42023418076), and individual study data were extracted for statistical re-analysis.

Results:

Twenty overlapping systematic reviews, 7, 17, and 8 references from RCTs, cohort studies, and joint replacement registries form the basis of this work. According to current best evidence, it is (i) 15-33 times more likely that 4th-generation alumina-zirconia pairings avoid a revision for infection than causing a revision for audible noise, (ii) 38-85 times more likely that 4th-generation alumina-zirconia pairings avoid a revision for infection than causing a revision for ceramic head fractures, and (iii) three to six times more likely that 4th-generation alumina-zirconia pairings avoid a revision for infection than cause a revision for ceramic liner fractures.

Conclusion:

Fourth-generation alumina-zirconia pairings in THA show a favorable benefit-risk ratio, with rare compound-specific adverse events and complications significantly outbalanced by long-term advantages, such as a markedly lower incidence of revision for infection.
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: EFORT Open Rev Año: 2024 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Idioma: En Revista: EFORT Open Rev Año: 2024 Tipo del documento: Article País de afiliación: Alemania
...