Your browser doesn't support javascript.
loading
In vitro evaluation of flow diverter performance using a human fibrinogen-based flow model.
Bilgin, Cem; Bayraktar, Esref Alperen; Oliver, Alexander A; Li, Jiahui; Cebral, Juan R; Kallmes, David F; Kadirvel, Ramanathan.
Afiliación
  • Bilgin C; Departments of1Radiology and.
  • Bayraktar EA; Departments of1Radiology and.
  • Oliver AA; Departments of1Radiology and.
  • Li J; Departments of1Radiology and.
  • Cebral JR; 2Center for Computational Fluid Dynamics, George Mason University, Fairfax, Virginia.
  • Kallmes DF; Departments of1Radiology and.
  • Kadirvel R; Departments of1Radiology and.
J Neurosurg ; : 1-8, 2024 Jul 12.
Article en En | MEDLINE | ID: mdl-38996402
ABSTRACT

OBJECTIVE:

Fibrin deposition represents a key step in aneurysm occlusion, promoting endothelization of implants and connective tissue organization as part of the aneurysm-healing mechanism. In this study, the authors introduce a novel in vitro testing platform for flow diverters based on human fibrinogen.

METHODS:

A flow diverter was deployed in 4 different glass models. The glass models had the same internal parent artery (4 mm) and aneurysm (8 mm) diameters with varying parent artery angulations (paraophthalmic, sidewall, bifurcation, and slightly curved models). The neck size and area were 4 mm and 25 mm2, respectively. Human fibrinogen (330 mg/dl) was circulated within the glass models at varying flow rates (0, 3, 4, and 5 ml/sec) with or without heparin, calcium chloride, and thrombin for as long as 6 hours or until complete fibrin coverage of the flow diverter's neck was achieved. Aneurysm neck coverage was defined as macroscopic fibrin deposition occluding the flow diverters' pores. Flow characteristics after flow diverter deployment were assessed with computational fluid dynamics analysis. The effects of flow rates, heparin, calcium chloride, and thrombin on fibrin deposition rates were tested using 1-way ANOVA and the Tukey test.

RESULTS:

A total of 84 replicates were performed. Human fibrin did not accumulate on the flow diverter stents under static conditions. The fibrin deposition rate on the aneurysm neck was significantly greater with the 5 ml/sec flow rate as compared to 3 ml/sec for all models. The paraophthalmic model had the highest inflow velocity of 48.7 cm/sec. The bifurcation model had the highest maximum shear stress (SS) and maximum normalized shear stress values at the device cells at 843.3 dyne/cm2 and 35.1 SS/SSinflow, respectively. The fibrin deposition rates of the paraophthalmic and bifurcation models were significantly higher than those of sidewall and slightly curved models for all additive or flow rate comparisons (p = 0.001 for all comparisons). The incorporation of thrombin significantly increased the fibrin deposition rates across all models (p = 0.001 for all models).

CONCLUSIONS:

Rates of fibrin deposition varied widely across different configurations and additive conditions in this novel in vitro model system. Fibrin accumulation started at the aneurysm inflow zone where flow velocity and shear stress were the highest. The primary factors influencing fibrin deposition included flow velocities, shear stress, and the addition of thrombin at a physiological concentration. Further research is needed to test the clinical utility of fibrinogen-based models for patient-specific aneurysms.