Influence of Gadolinium-based Contrast Media and Inter-reader Variation on the Estimation of Intravoxel Incoherent Motion (IVIM) Parameters in Breast MR Imaging.
Fueger, Barbara J; Varga, Raoul; Kapetas, Panagiotis; Pötsch, Nina; Helbich, Thomas H; Baltzer, Pascal A T; Clauser, Paola.
Afiliación
  • Fueger BJ; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
  • Varga R; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
  • Kapetas P; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
  • Pötsch N; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
  • Helbich TH; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
  • Baltzer PAT; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
  • Clauser P; Department of Biomedical Imaging and Image-guided Biopsy, Division of Molecular and Gender Imaging, Medical University of Vienna, Vienna, Austria.
Magn Reson Med Sci ; 2024 Jul 13.
Article en En | MEDLINE | ID: mdl-39010211
ABSTRACT

PURPOSE:

Gadolinium-based contrast media (GBCM) may affect apparent diffusion coefficient measurements on diffusion-weighted imaging. We aimed at investigating the effect of GBCM and inter-reader variation on intravoxel incoherent motion (IVIM) parameters in breast lesions.

METHODS:

A total of 89 patients referred to 3T breast MRI with at least one histologically verified lesion were included. IVIM data were acquired using a single-shot echo planar imaging sequence before and after GBCM administration. D (true diffusion coefficient), D* (pseudo-diffusion coefficient) and f (perfusion fraction) were calculated and measured by two readers (R1, R2). Inter-reader and intra-reader agreements were assessed by intraclass correlation coefficients (ICCs) and Bland-Altman plots.

RESULTS:

D was comparable before and after GBCM administration and between readers. D* and f decreased after GBCM administration and showed a lower agreement between readers. Intra-reader agreement before and after GBCM administration was almost perfect for D for both R1 and R2 (ICC 0.955 and 0.887). The intra-reader agreement was substantial to moderate for D* (ICC R1 0.708, R2 0.583) and moderate for f (ICC R1 0.529 and R2 0.425). Inter-reader agreement before GBCM administration was almost perfect for D (ICC 0.905), substantial for D* (ICC 0.733), and moderate for f (ICC 0.404); after contrast media administration, it was almost perfect for D (ICC 0.876) and substantial for D* (ICC 0.654) and f (ICC 0.606). Bland-Altman plots revealed no significant bias.

CONCLUSION:

Administration of GBCM seems to have a stronger effect on D* and f values than on D values. This should be considered when applying IVIM in clinical practice.