Improving the Sensitivity of Protein Quantification by Immunoaffinity Liquid Chromatography─Triple Quadrupole Mass Spectrometry Using an Iterative Transition Summing Technique.
Johnson, Jay S; Palandra, Joe; Psychogios, Nikolaos; Walsh, Jason M; Neubert, Hendrik.
Afiliación
  • Johnson JS; Pharmacokinetics, Dynamics & Metabolism (PDM), Pfizer, Andover, Massachusetts 01810, United States.
  • Palandra J; Pharmacokinetics, Dynamics & Metabolism (PDM), Pfizer, Andover, Massachusetts 01810, United States.
  • Psychogios N; Pharmacokinetics, Dynamics & Metabolism (PDM), Pfizer, Andover, Massachusetts 01810, United States.
  • Walsh JM; Pharmacokinetics, Dynamics & Metabolism (PDM), Pfizer, Andover, Massachusetts 01810, United States.
  • Neubert H; Pharmacokinetics, Dynamics & Metabolism (PDM), Pfizer, Andover, Massachusetts 01810, United States.
Anal Chem ; 96(36): 14326-14331, 2024 Sep 10.
Article en En | MEDLINE | ID: mdl-39186623
ABSTRACT
The desire to reach ever-diminishing lower limits of quantification (LLOQ) to probe changes in low abundance protein targets has led to enormous progress in sample preparation and liquid chromatography-tandem mass spectrometry (LC-MS/MS) instrumentation. To maximize signal and reduce noise, many approaches have been employed, including specific immunoaffinity (IA) enrichment and reducing the LC flow to the nanoflow (nLC) level; however, additional sensitivity gains may still be required. Recently, a technique termed "echo summing" has been described for small-molecular-weight analytes on a triple quadrupole (QqQ) MS where multiple iterations of the same, single selected reaction monitoring (SRM) transition are collected, summed, and integrated, yielding significant analyte dependent signal-to-noise (S/N) improvements. Herein, the direct applicability of echo summing to protein quantification by sequential IA combined with nLC-MS/MS (IA-nLC-MS/MS) is described for a beta nerve growth factor (NGF) and a soluble asialoglycoprotein receptor (sASGPR) assay from human serum. Five iterations of echo summing outperformed traditional collection in relative average accuracy (-1.5 ± 7.7 vs -41.7 ± 10.7% bias) and precision (7.8 vs 18.4% coefficient of variation (CV)) of the low-end quality control (QC) sample (N = 4) for NGF and improved functional sensitivity of serially diluted serum QC samples (N = 5 each population) approximately 2-fold (1.96 and 2.00-fold) for two peptides of sASGPR. Echo summing also extended the minimum quantifiable QC level for sASGPR 4-fold lower. Similar gains are believed to be achievable for most protein IA-nLC-MS/MS assays.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Espectrometría de Masas en Tándem Límite: Humans Idioma: En Revista: Anal Chem Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos