Development of a µPAD for the point-of-care testing of serum glutamic oxaloacetic transaminase (SGOT).
Resmi, P E; Aarathi, Pradeep; Suneesh, P V; Ramachandran, T; Bipin, G Nair; Satheesh, Babu T G.
Afiliación
  • Resmi PE; Department of Chemistry, Amrita School of Physical Sciences Coimbatore, Amrita Vishwa Vidyapeetham, Coimbatore, 641112, India.
  • Aarathi P; Amrita Biosensor Research Lab, Amrita School of Engineering Coimbatore, Amrita Vishwa Vidyapeetham, Coimbatore, 641112, India.
  • Suneesh PV; Department of Chemistry, Amrita School of Physical Sciences Coimbatore, Amrita Vishwa Vidyapeetham, Coimbatore, 641112, India.
  • Ramachandran T; Amrita Biosensor Research Lab, Amrita School of Engineering Coimbatore, Amrita Vishwa Vidyapeetham, Coimbatore, 641112, India.
  • Bipin GN; Department of Chemistry, Amrita School of Physical Sciences Coimbatore, Amrita Vishwa Vidyapeetham, Coimbatore, 641112, India.
  • Satheesh BTG; Amrita Biosensor Research Lab, Amrita School of Engineering Coimbatore, Amrita Vishwa Vidyapeetham, Coimbatore, 641112, India.
Mikrochim Acta ; 191(10): 619, 2024 09 25.
Article en En | MEDLINE | ID: mdl-39320528
ABSTRACT
A wax-patterned paper analytical device (µPAD) has been developed for point-of-care colourimetric testing of serum glutamic oxaloacetic transaminase (SGOT). The detection method was based on the transamination reaction of aspartate with α-ketoglutarate, leading to the formation of oxaloacetate which reacts with the reagent Fast Blue BB salt and forms a cavern pink colour. The intensity of the cavern pink colour grows as the concentration of SGOT increases. UV-visible spectroscopy was utilized to optimize reaction conditions, and the optimized reagents were dropped onto the wax-patterned paper. The coloured PADs, after the addition of SGOT, have been photographed, and a colour band has been generated to correlate the SGOT concentration visually. The images were used to calculate the intensity values using ImageJ software, which inturn was used to calculate the SGOT concentration. The PADs were also tested with serum samples, and SGOT spiked serum samples. The PAD could detect the SGOT concentration ranging from 5 to 200 U/L. The analysis yielded highly accurate results with less than 6% relative error compared to the clinical sample. This colourimetric test demonstrated exceptional selectivity in the presence of other biomolecules in the blood serum, with a detection limit of 2.77 U/L and a limit of quantification of 9.25 U/L. Additionally, a plasma separation membrane was integrated with the PAD to directly test SGOT from finger-prick blood samples.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Aspartato Aminotransferasas / Colorimetría / Pruebas en el Punto de Atención Límite: Humans Idioma: En Revista: Mikrochim Acta Año: 2024 Tipo del documento: Article País de afiliación: India

Texto completo: 1 Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Aspartato Aminotransferasas / Colorimetría / Pruebas en el Punto de Atención Límite: Humans Idioma: En Revista: Mikrochim Acta Año: 2024 Tipo del documento: Article País de afiliación: India
...