Your browser doesn't support javascript.
loading
Simple reversed-phase high-performance liquid chromatography quantitation of ganciclovir in human serum and urine.
Page, T; Sherwood, C; Connor, J D; Tarnowski, T.
Afiliación
  • Page T; Department of Neurosciences, University of California, San Diego, La Jolla 92093, USA.
J Chromatogr B Biomed Appl ; 675(2): 342-6, 1996 Jan 26.
Article en En | MEDLINE | ID: mdl-8852725
A fast, simple, and cost-effective HPLC method for the quantitation of the antiviral drug ganciclovir is described. The serum samples are extracted with perchloric acid and neutralized with potassium phosphate buffer, and urine samples are diluted with distilled water. A reversed-phase column with isocratic elution by 15 mM potassium phosphate buffer (pH 2.5) containing 0.25% acetonitrile is used to separate ganciclovir; quantitation is by UV absorbance at 254 nm. Total turnaround time is 22 min; more than 3000 samples can be run on a single column without loss of peak quality. The limit of quantitation is 0.05 micrograms/ml. Recoveries varied from 91 to 107% with coefficients of variation ranging from 0.387 to 7.95%.
Asunto(s)
Buscar en Google
Colección: 01-internacional Base de datos: MEDLINE Asunto principal: Antivirales / Ganciclovir / Cromatografía Líquida de Alta Presión Límite: Humans Idioma: En Revista: J Chromatogr B Biomed Appl Año: 1996 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google