Ion fragmentation in an electrospray ionization mass spectrometer interface with different gases.
Rapid Commun Mass Spectrom
; 15(4): 249-57, 2001.
Article
em En
| MEDLINE
| ID: mdl-11223955
Nitrogen, argon, and krypton are used as curtain gases in an electrospray ionization mass spectrometer in an attempt to study the effect of these gases on the extent of ion fragmentation between the orifice and the skimmer of the interface region. A previously published collision model predicts that the degree of ion fragmentation increases with increasing mass of the curtain gas. However, the fragmentation yields are found to be the opposite to that expected. It is believed that the reversed trend with argon and krypton is caused by condensation of the gases within the free jet expansion between the orifice and the skimmer. A condensation parameter can be used to predict the degree of clustering of gases within a free jet expansion. When the condensation parameter is minimized, the predicted trend of fragmentation with mass is observed.