Functional imaging of parotid glands: diffusion-weighted echo-planar MRI before and after stimulation.
Habermann, C R; Cramer, M C; Graessner, J; Gossrau, P; Reitmeier, F; Fiehler, J; Schoder, V; Jaehne, M; Adam, G.
Afiliação
  • Habermann CR; Department of Diagnostic and Interventional Radiology, University Hospital Hamburg-Eppendorf, Hamburg, Germany. c.habermann@uke.uni-hamburg.de
Rofo ; 176(10): 1385-9, 2004 Oct.
Article em En | MEDLINE | ID: mdl-15383968
ABSTRACT

PURPOSE:

To investigate the feasibility of diffusion-weighted (DW) echo-planar imaging (EPI) for measuring different functional conditions of the parotid gland and to compare different measurement approaches. MATERIALS AND

METHODS:

Parotid glands of 27 healthy volunteers were examined with a DW EPI sequence (TR 1,500 msec, TE 77 msec, field-of-view 250 x 250 mm, pixel size 2.10 x 1.95 mm, section thickness 5 mm) before and after oral stimulation with commercially available lemon juice. The b factors used were 0, 500, and 1,000 sec/mm (2). Apparent diffusion coefficient (ADC) maps were digitally transferred to MRIcro (Chris Rorden, University of Nottingham, Great Britain) and evaluated with a manually placed circular region of interest (ROI) containing 100 - 200 pixel. Additional ROIs including the entire parotid gland were placed on either side. The results of both measurements were compared, using the Student's t test based on the median ADC values for each person. A two-tailed p-value of less than.05 was determined to indicate statistical significance. To compare both measurement approaches, the Pearson's correlation coefficient (r) was calculated.

RESULTS:

Diffusion-weighted echo-planar MR imaging successfully visualized the parotid glands of all volunteers. In a first step, the median ADC value per person was computed. Using ROIs of 100 - 200 pixels, the mean was calculated to be 1.08 x 10 (- 3) mm (2)/sec +/- 0.12 x 10 (- 3) mm (2)/sec for both parotid glands prior to stimulation. After stimulation, the mean ADC was measured at 1.15 x 10 (- 3) mm (2)/sec +/- 0.11 x 10 (- 3) mm (2)/sec for both parotid glands. Evaluating the entire parotid gland, the ADC was 1.12 x 10 (- 3) mm (2)/sec +/- 0.08 x 10 (- 3) mm (2)/sec prior to stimulation, whereas the ADC increased to 1.18 x 10 (- 3) mm (2)/sec +/- 0.09 x 10 (- 3) mm (2)/sec after stimulation with lemon juice. For both types of measurements, the increase in ADC after stimulation proved to be significant (p < 0.001). High correlations between both measurement types were found (r >.83).

CONCLUSION:

Diffusion-weighted echo-planar MR imaging allows non-invasive quantification of functional changes in the parotid glands.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glândula Parótida / Imagem Ecoplanar Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Rofo Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Alemanha
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Glândula Parótida / Imagem Ecoplanar Tipo de estudo: Evaluation_studies / Prognostic_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Rofo Ano de publicação: 2004 Tipo de documento: Article País de afiliação: Alemanha
...