Measurement of total and differential white blood cell counts in synovial fluid by means of an automated hematology analyzer.
Sugiuchi, Hiroyuki; Ando, Yukio; Manabe, Masahiro; Nakamura, Eiichi; Mizuta, Hiroshi; Nagata, Shiro; Okabe, Hiroaki.
Afiliação
  • Sugiuchi H; Department of Laboratory Medicine, Kumamoto University Medical School, Kumamoto, Japan.
J Lab Clin Med ; 146(1): 36-42, 2005 Jul.
Article em En | MEDLINE | ID: mdl-16025090
ABSTRACT
We developed a rapid and accurate method for quantifying total and differential white blood cell (WBC) counts by pretreating synovial fluid with hyaluronidase and using an automated hematology analyzer. Forty-seven samples of synovial fluid that had been placed in blood-collection tubes containing ethylenediamine- N,N,N',N' -tetraacetic acid as an anticoagulant were treated with hyaluronidase at 37 degrees C for 10 minutes, and then the total and differential WBC counts were determined by means of the automated hematology analyzer. Results were compared with those achieved by traditional manual counting methods. For the automated method, the coefficient of variation values for within-run precision of the WBC count were 5.27%, 3.56%, and 3.01% at 0.54, 1.12, and 2.05 x 10(9) /L, respectively; the run-to-run coefficient of variation values was less than 10.0%. The total and differential WBC counts obtained by this automated method showed good correlation with those obtained by the hemocytometer method ( r = .998; P < .0001; regression formula, y = 0.986 x - 0.072). Bland-Altman plots indicated no significant discrepancy between the methods. Our evaluation supports the use of this automated hematology analyzer method to measure total and differential WBC counts, which should aid clinical diagnosis.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Líquido Sinovial / Contagem de Leucócitos Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: J Lab Clin Med Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Japão
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Líquido Sinovial / Contagem de Leucócitos Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: J Lab Clin Med Ano de publicação: 2005 Tipo de documento: Article País de afiliação: Japão
...