Phase-contrast velocimetry with hyperpolarized 3He for in vitro and in vivo characterization of airflow.
Magn Reson Med
; 55(6): 1318-25, 2006 Jun.
Article
em En
| MEDLINE
| ID: mdl-16700024
This paper describes a technique that combines radial MRI and phase contrast (PC) to map the velocities of hyperpolarized gases ((3)He) in respiratory airways. The method was evaluated on well known geometries (straight and U-shaped pipes) before it was applied in vivo. Dynamic 2D maps of the three velocity components were obtained from a 10-mm slice with an in-plane spatial resolution of 1.6 mm within 1 s. Integration of the in vitro through-plane velocity over the slice matched the input flow within a relative precision of 6.4%. As expected for the given Reynolds number, a parabolic velocity profile was obtained in the straight pipe. In the U-shaped pipe the three velocity components were measured and compared to a fluid-dynamics simulation so the precision was evaluated as fine as 0.025 m s(-1). The technique also demonstrated its ability to visualize vortices and localize characteristic points, such as the maximum velocity and vortex-center positions. Finally, in vivo feasibility was demonstrated in the human trachea during inhalation.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Reologia
/
Imageamento por Ressonância Magnética
/
Interpretação de Imagem Assistida por Computador
/
Aumento da Imagem
/
Ventilação Pulmonar
/
Hélio
/
Pulmão
Limite:
Humans
Idioma:
En
Revista:
Magn Reson Med
Assunto da revista:
DIAGNOSTICO POR IMAGEM
Ano de publicação:
2006
Tipo de documento:
Article
País de afiliação:
França