Microdialysis-based 48-hour continuous glucose monitoring with GlucoDay: clinical performance and patients' acceptance.
Diabetes Technol Ther
; 8(5): 570-5, 2006 Oct.
Article
em En
| MEDLINE
| ID: mdl-17037971
ABSTRACT
BACKGROUND:
This study was designed to assess clinical performance and patients' acceptance of the minimally invasive microdialysis-based continuous glucose monitoring system Gluco- Day() (Menarini Diagnostics, Florence, Italy) with a targeted monitoring time of 48 h.METHODS:
An inpatient sample of 28 patients with diabetes was studied. The analysis of clinical performance was performed using mean absolute differences (MAD) (in percent), Pearson correlations, the Bland-Altman analysis, and Clarke Error Grid Analysis (EGA). GlucoDay glucose values were compared with laboratory standard blood glucose measurements (glucohexokinase assay). The patients' acceptance of the monitoring device was assessed via two self-report scales (pain during application and discomfort while wearing device).RESULTS:
A mean (+/- SD) monitoring time of 45.7 +/- 3.3 h with a total of 484 paired readings could be achieved. A correlation of r (average) = 0.91 and a MAD of 19.9% indicated satisfactory to good clinical performance. Of the paired readings, 95.5% fell into the acceptable A and B zones of the EGA. Rather wide 95% limits of agreement were revealed in the Bland-Altman analysis. Whereas virtually no pain was experienced during sensor application, discomfort associated with wearing the device was rather high. All of the participants, however, stated that they would wear the device again.CONCLUSIONS:
Satisfactory to good performance of the GlucoDay monitor was observed, indicating the device to be suitable for routine clinical use. In particular, however, the discomfort experienced during wearing requires further improvements in its usability.
Buscar no Google
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Glicemia
/
Técnicas Biossensoriais
/
Automonitorização da Glicemia
/
Satisfação do Paciente
/
Microdiálise
/
Diabetes Mellitus Tipo 1
Limite:
Adult
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Diabetes Technol Ther
Assunto da revista:
ENDOCRINOLOGIA
/
TERAPEUTICA
Ano de publicação:
2006
Tipo de documento:
Article
País de afiliação: