Your browser doesn't support javascript.
loading
Design and performance of a respiratory amplitude gating device for PET/CT imaging.
Chang, Guoping; Chang, Tingting; Clark, John W; Mawlawi, Osama R.
Afiliação
  • Chang G; Electrical and Computer Engineering, Rice University, Houston, Texas 77005, USA.
Med Phys ; 37(4): 1408-12, 2010 Apr.
Article em En | MEDLINE | ID: mdl-20443462
ABSTRACT

PURPOSE:

Recently, the authors proposed a free-breathing amplitude gating (FBAG) technique for PET/CT scanners. The implementation of this technique required specialized hardware and software components that were specifically designed to interface with commercial respiratory gating devices to generate the necessary triggers required for the FBAG technique. The objective of this technical note is to introduce an in-house device that integrates all the necessary hardware and software components as well as tracks the patient's respiratory motion to realize amplitude gating on PET/CT scanners.

METHODS:

The in-house device is composed of a piezoelectric transducer coupled to a data-acquisition system in order to monitor the respiratory waveform. A LABVIEW program was designed to control the data-acquisition device and inject triggers into the PET list stream whenever the detected respiratory amplitude crossed a predetermined amplitude range. A timer was also programmed to stop the scan when the accumulated time within the selected amplitude range REACHED a user-set interval. This device was tested using a volunteer and a phantom study.

RESULTS:

The results from the volunteer and phantom studies showed that the in-house device can detect similar respiratory signals as commercially available respiratory gating systems and is able to generate the necessary triggers to suppress respiratory motion artifacts.

CONCLUSIONS:

The proposed in-house device can be used to implement the FBAG technique in current PET/CT scanners.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Diagnóstico por Imagem / Tomografia Computadorizada por Raios X / Tomografia por Emissão de Pósitrons Tipo de estudo: Diagnostic_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Med Phys Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Diagnóstico por Imagem / Tomografia Computadorizada por Raios X / Tomografia por Emissão de Pósitrons Tipo de estudo: Diagnostic_studies Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Med Phys Ano de publicação: 2010 Tipo de documento: Article País de afiliação: Estados Unidos
...