An upgraded x-ray spectroscopy diagnostic on MST.
Rev Sci Instrum
; 81(10): 10E308, 2010 Oct.
Article
em En
| MEDLINE
| ID: mdl-21034007
An upgraded x-ray spectroscopy diagnostic is used to measure the distribution of fast electrons in MST and to determine Z(eff) and the particle diffusion coefficient D(r). A radial array of 12 CdZnTe hard-x-ray detectors measures 10-150 keV Bremsstrahlung from fast electrons, a signature of reduced stochasticity and improved confinement in the plasma. A new Si soft-x-ray detector measures 2-10 keV Bremsstrahlung from thermal and fast electrons. The shaped output pulses from both detector types are digitized and the resulting waveforms are fit with Gaussians to resolve pileup and provide good time and energy resolution. Lead apertures prevent detector saturation and provide a well-known etendue, while lead shielding prevents pickup from stray x-rays. New Be vacuum windows transmit >2 keV x-rays, and additional Al and Be filters are sometimes used to reduce low energy flux for better resolution at higher energies. Measured spectra are compared to those predicted by the Fokker-Planck code CQL3D to deduce Z(eff) and D(r).