Efficacy of two commercially available cardiac pacing catheters for transesophageal atrial pacing in dogs.
Chapel, Emily H; Sanders, Robert A.
Afiliação
  • Chapel EH; Michigan State University, College of Veterinary Medicine, East Lansing, MI 48823, USA.
J Vet Cardiol ; 14(3): 409-14, 2012 Sep.
Article em En | MEDLINE | ID: mdl-22841903
ABSTRACT

OBJECTIVE:

Compare the efficacy of two cardiac pacing catheters for transesophageal atrial pacing (TAP). ANIMALS Ten healthy dogs.

METHODS:

Transesophageal atrial pacing was attempted in left lateral recumbency under general anesthesia. In series, a curved electrophysiologic (EP) catheter and a straight transesophageal pacing (TP) catheter were passed transorally into the distal esophagus, caudal to the heart. Each catheter was slowly withdrawn until pacing was initiated. The catheter continued to be withdrawn in 5 mm increments until capture was no longer achieved at maximal pacing setting creating a total zone of capture (tZOC). Minimum pacing threshold (Th(min)) and degree of extraneous muscle stimulation (EMS) were determined at each site.

RESULTS:

The EP catheter achieved TAP in 10/10 dogs with a Th(min) of 8.8 ± 4.8 mA and tZOC of 4.0 ± 1.7 cm. The TP catheter achieved TAP in 7/10 dogs with a Th(min) of 22.5 ± 5.6 mA and tZOC of 1.2 ± 1.6 cm. No EMS was noted during TAP at Th(min) using the EP catheter. Of the 7 dogs in which TAP was successful, 3/7 experienced considerable EMS at Th(min). There was no significant relationship between patient body weight and Th(min) with either catheter (EP catheter P = 0.09; TP catheter P = 0.81).

CONCLUSIONS:

The electrophysiology catheter captured the atria of more dogs over a larger area without EMS as compared to the TP catheter. The electrophysiology catheter is recommended for transesophageal atrial pacing in dogs.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Estimulação Cardíaca Artificial / Cães / Esôfago Limite: Animals Idioma: En Revista: J Vet Cardiol Assunto da revista: CARDIOLOGIA / MEDICINA VETERINARIA Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Marca-Passo Artificial / Estimulação Cardíaca Artificial / Cães / Esôfago Limite: Animals Idioma: En Revista: J Vet Cardiol Assunto da revista: CARDIOLOGIA / MEDICINA VETERINARIA Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Estados Unidos
...