Your browser doesn't support javascript.
loading
Comparison of CT Fluoroscopy-Guided Manual and CT-Guided Robotic Positioning System for In Vivo Needle Placements in Swine Liver.
Cornelis, F; Takaki, H; Laskhmanan, M; Durack, J C; Erinjeri, J P; Getrajdman, G I; Maybody, M; Sofocleous, C T; Solomon, S B; Srimathveeravalli, G.
Afiliação
  • Cornelis F; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Takaki H; Department of Radiology, Pellegrin Hospital, Place Amélie Raba Léon, 33076, Bordeaux, France.
  • Laskhmanan M; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Durack JC; Perfint Healthcare Inc, Chennai, Tamil Nadu, India.
  • Erinjeri JP; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Getrajdman GI; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Maybody M; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Sofocleous CT; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Solomon SB; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
  • Srimathveeravalli G; Interventional Radiology Service, Department of Radiology, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, NY, 10065, USA.
Cardiovasc Intervent Radiol ; 38(5): 1252-60, 2015 Oct.
Article em En | MEDLINE | ID: mdl-25376924
ABSTRACT

PURPOSE:

To compare CT fluoroscopy-guided manual and CT-guided robotic positioning system (RPS)-assisted needle placement by experienced IR physicians to targets in swine liver. MATERIALS AND

METHODS:

Manual and RPS-assisted needle placement was performed by six experienced IR physicians to four 5 mm fiducial seeds placed in swine liver (n = 6). Placement performance was assessed for placement accuracy, procedure time, number of confirmatory scans, needle manipulations, and procedure radiation dose. Intra-modality difference in performance for each physician was assessed using paired t test. Inter-physician performance variation for each modality was analyzed using Kruskal-Wallis test.

RESULTS:

Paired comparison of manual and RPS-assisted placements to a target by the same physician indicated accuracy outcomes was not statistically different (manual 4.53 mm; RPS 4.66 mm; p = 0.41), but manual placement resulted in higher total radiation dose (manual 1075.77 mGy/cm; RPS 636.4 mGy/cm; p = 0.03), required more confirmation scans (manual 6.6; RPS 1.6; p < 0.0001) and needle manipulations (manual 4.6; RPS 0.4; p < 0.0001). Procedure time for RPS was longer than manual placement (manual 6.12 min; RPS 9.7 min; p = 0.0003). Comparison of inter-physician performance during manual placement indicated significant differences in the time taken to complete placements (p = 0.008) and number of repositions (p = 0.04) but not in other study measures (p > 0.05). Comparison of inter-physician performance during RPS-assisted placement suggested statistically significant differences in procedure time (p = 0.02) and not in other study measures (p > 0.05).

CONCLUSIONS:

CT-guided RPS-assisted needle placement reduced radiation dose, number of confirmatory scans, and needle manipulations when compared to manual needle placement by experienced IR physicians, with equivalent accuracy.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Robótica / Tomografia Computadorizada por Raios X / Radiografia Intervencionista / Cirurgia Assistida por Computador / Fígado / Agulhas Limite: Animals Idioma: En Revista: Cardiovasc Intervent Radiol Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Robótica / Tomografia Computadorizada por Raios X / Radiografia Intervencionista / Cirurgia Assistida por Computador / Fígado / Agulhas Limite: Animals Idioma: En Revista: Cardiovasc Intervent Radiol Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Estados Unidos
...