Cost of a recall of a single-center experience managing the Riata defibrillator lead.
Am J Cardiol
; 115(2): 206-8, 2015 Jan 15.
Article
em En
| MEDLINE
| ID: mdl-25479754
Riata and Riata ST defibrillator leads (St. Jude Medical, Sylmar, California) were recalled in 2011 due to increased risk of insulation failure leading to externalized cables. Fluoroscopic screening can identify insulation failure, although the relation between mechanical failure and electrical failure is unclear. At the time of the recall, the University of Virginia developed a screening program, including fluoroscopic evaluation, education sessions, device interrogation, and remote monitoring for patients with this defibrillator lead. The aim of this study was to review the outcomes of the screening program, including costs, which were absorbed by our institution. Costs were calculated using Medicare reimbursement estimates. Forty-eight patients participated in the screening program. At initial screening, 31% were found to have evidence of insulation failure but electrical function was normal in all leads. The cost of this program was $35,358.72. The cost per diagnosis of mechanical lead failure was $2,357.25. During 2 years of follow-up, 1 patient experienced Riata lead electrical failure without fluoroscopic evidence of insulation failure. Patients were more likely to have a lead revision if there was evidence of insulation failure. Lead revisions occurred at the time of generator change in 88% of patients with insulation failure but in only 14% of patients with a fluoroscopically normal lead (p = 0.04). The cost of recall-related defibrillator lead revisions was $81,704.55. In conclusion, our Riata screening program added expense without clear benefit to patients. In fact, patients may have been put at more risk by undergoing defibrillator lead revisions based solely on the results of the fluoroscopic screening.
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Contexto em Saúde:
1_ASSA2030
Problema de saúde:
1_financiamento_saude
Assunto principal:
Desfibriladores Implantáveis
/
Recall de Dispositivo Médico
Tipo de estudo:
Health_economic_evaluation
/
Observational_studies
/
Risk_factors_studies
Limite:
Female
/
Humans
/
Male
País/Região como assunto:
America do norte
Idioma:
En
Revista:
Am J Cardiol
Ano de publicação:
2015
Tipo de documento:
Article