Your browser doesn't support javascript.
loading
Highly sensitive and novel point-of-care system, aQcare Chlamydia TRF kit for detecting Chlamydia trachomatis by using europium (Eu) (III) chelated nanoparticles.
Ham, Ji Yeon; Jung, Jaean; Hwang, Byung-Gap; Kim, Won-Jung; Kim, Young-Seop; Kim, Eun-Ju; Cho, Mi-Yeon; Hwang, Mi-Sun; Won, Dong Il; Suh, Jang Soo.
Afiliação
  • Ham JY; Department of Laboratory Medicine, Kyungpook National University Hospital, Daegu, Korea.
  • Jung J; Medisensor Inc., Daegu, Korea.
  • Hwang BG; Medisensor Inc., Daegu, Korea.
  • Kim WJ; Medisensor Inc., Daegu, Korea.
  • Kim YS; Medisensor Inc., Daegu, Korea.
  • Kim EJ; Medisensor Inc., Daegu, Korea.
  • Cho MY; Medisensor Inc., Daegu, Korea.
  • Hwang MS; Medisensor Inc., Daegu, Korea.
  • Won DI; Department of Laboratory Medicine, Kyungpook National University Hospital, Daegu, Korea. ; Department of Clinical Pathology, Kyungpook National University School of Medicine, Daegu, Korea.
  • Suh JS; Department of Laboratory Medicine, Kyungpook National University Hospital, Daegu, Korea. ; Department of Clinical Pathology, Kyungpook National University School of Medicine, Daegu, Korea.
Ann Lab Med ; 35(1): 50-6, 2015 Jan.
Article em En | MEDLINE | ID: mdl-25553280
ABSTRACT

BACKGROUND:

The bacterium Chlamydia trachomatis is one of the leading causes of sexually transmitted diseases worldwide. Since no simple and effective tool exists to diagnose C. trachomatis infections, we evaluated a novel point-of-care (POC) test, aQcare Chlamydia TRF kit, which uses europium-chelated nanoparticles and a time-resolved fluorescence reader.

METHODS:

The test performance was evaluated by comparing the results obtained using the novel POC testing kit with those obtained using a nucleic acid amplification test (NAAT), using 114 NAAT-positive and 327 NAAT-negative samples.

RESULTS:

The cut-off value of the novel test was 20.8 with a detection limit of 0.27 ng/mL. No interference or cross-reactivity was observed. Diagnostic accuracy showed an overall sensitivity of 93.0% (106/114), specificity of 96.3% (315/327), positive predictive value (PPV) of 89.8% (106/118), and negative predictive value (NPV) of 97.5% (315/323). The sensitivity of the novel test was much higher than that of currently available POC tests. Furthermore, the relative ease and short turnaround time (30 min) of this assay enables C. trachomatis-infected individuals to be treated without a diagnostic delay.

CONCLUSIONS:

This simple and novel test is a potential tool to screen a larger population, especially those in areas with limited resources.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções por Chlamydia / Chlamydia trachomatis / Európio / Nanopartículas Metálicas Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Ann Lab Med Ano de publicação: 2015 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Infecções por Chlamydia / Chlamydia trachomatis / Európio / Nanopartículas Metálicas Tipo de estudo: Diagnostic_studies / Evaluation_studies / Prognostic_studies Limite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Ann Lab Med Ano de publicação: 2015 Tipo de documento: Article
...