Your browser doesn't support javascript.
loading
Free breathing three-dimensional late gadolinium enhancement cardiovascular magnetic resonance using outer volume suppressed projection navigators.
Menon, Rajiv G; Miller, G Wilson; Jeudy, Jean; Rajagopalan, Sanjay; Shin, Taehoon.
Afiliação
  • Menon RG; Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, Baltimore, Maryland, USA.
  • Miller GW; Department of Radiology & Medical Imaging, University of Virginia School of Medicine, Charlottesville, Virginia, USA.
  • Jeudy J; Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, Baltimore, Maryland, USA.
  • Rajagopalan S; Division of Cardiovascular Medicine, University of Maryland, Baltimore, Maryland, USA.
  • Shin T; Department of Diagnostic Radiology and Nuclear Medicine, University of Maryland, Baltimore, Maryland, USA.
Magn Reson Med ; 77(4): 1533-1543, 2017 04.
Article em En | MEDLINE | ID: mdl-27122450
ABSTRACT

PURPOSE:

To develop a three-dimensional, free-breathing, late gadolinium enhancement (3D FB-LGE) cardiovascular magnetic resonance (CMR) technique, and to compare it with clinically used two-dimensional breath-hold LGE (2D BH-LGE).

METHODS:

The proposed 3D FB-LGE method consisted of inversion preparation, inversion delay, fat saturation, outer volume suppression, one-dimensional projection navigators, and a segmented stack of spirals acquisition. The 3D FB-LGE and 2D BH-LGE scans were performed on 29 cardiac patients. Qualitative analysis and quantitative analysis (in patients with scar) were performed.

RESULTS:

No significant differences were noted between the 3D FB-LGE and 2D BH-LGE data sets in terms of overall image quality score (2D 4.69 ± 0.60 versus 3D 4.55 ± 0.51, P = 0.46) and image artifact score (2D 1.10 ± 0.31 versus 3D 1.17 ± 0.38; P = 0.63). The average difference in fractional scar volume between the 3D and 2D methods was 1.9% (n = 5). Acquisition time was significantly shorter for the 3D FB-LGE over 2D BH-LGE by a factor of 2.83 ± 0.77 (P < 0.0001).

CONCLUSIONS:

The 3D FB-LGE is a viable option for patients, particularly in acute settings or in patients who are unable to comply with breath-hold instructions. Magn Reson Med 771533-1543, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Organometálicos / Aumento da Imagem / Miocárdio Atordoado / Imagem Cinética por Ressonância Magnética / Imageamento Tridimensional / Meglumina Tipo de estudo: Diagnostic_studies / Qualitative_research Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Compostos Organometálicos / Aumento da Imagem / Miocárdio Atordoado / Imagem Cinética por Ressonância Magnética / Imageamento Tridimensional / Meglumina Tipo de estudo: Diagnostic_studies / Qualitative_research Limite: Female / Humans / Male / Middle aged Idioma: En Revista: Magn Reson Med Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Estados Unidos
...