Your browser doesn't support javascript.
loading
Development of an Evaluation Device for Phagocytic Activity of New Phagocytes Using Simple and pH-sensitive Particles that Do Not Require Pre-treatment.
Zhang, Ran; Kobayashi, Yutaro; Kazumura, Kimiko; Tsuchiya, Hiroshi; Morishita, Naokazu; Inagawa, Hiroyuki; Soma, Gen-Ichiro.
Afiliação
  • Zhang R; Control of Innate Immunity, Technology Research Association, Kagawa, Japan zhang@shizenmeneki.org.
  • Kobayashi Y; Faculty of Medicine, Kagawa University, Kagawa, Japan.
  • Kazumura K; Central Research Laboratory, Hamamatsu Photonics K.K., Hamamatsu, Japan.
  • Tsuchiya H; Central Research Laboratory, Hamamatsu Photonics K.K., Hamamatsu, Japan.
  • Morishita N; Electron Tube Division, Hamamatsu Photonics K.K., Hamamatsu, Japan.
  • Inagawa H; Control of Innate Immunity, Technology Research Association, Kagawa, Japan Faculty of Medicine, Kagawa University, Kagawa, Japan Niigata University of Pharmacy and Applied Life Sciences, Niigata, Japan.
  • Soma G; Control of Innate Immunity, Technology Research Association, Kagawa, Japan Faculty of Medicine, Kagawa University, Kagawa, Japan Niigata University of Pharmacy and Applied Life Sciences, Niigata, Japan.
Anticancer Res ; 36(7): 3613-8, 2016 Jul.
Article em En | MEDLINE | ID: mdl-27354631
ABSTRACT
BACKGROUND/

AIM:

Phagocytic activity is affected by a number of different stress and age-dependent factors. An easy measurement of phagocytic activity is thought to allow an indicator of an individual's health. In this study, we investigated conditions of measurement to easily evaluate the activity of phagocytosis of phagocytic cells (macrophages and neutrophils) using an easy-to-use prototype, which was improved from the device by Hamamatsu Photonics K.K., to detect neutrophil activity using subtle fluorescence. MATERIALS AND

METHODS:

pH-sensitive fluorescent particles (pHrodo-Green E. coli Bio particles, GE particles) were added to mouse-derived macrophage cell lines (J774.1) and then incubated for 2 h at 37°C. For negative control, the phagocytosis inhibitor cytochalasin D (CyD), was added prior to culture. Next, fluorescence intensity was measured by the Prototype to evaluate the phagocytic activity of macrophages and neutrophils. Phagocytosis was also confirmed by flow cytometry.

RESULTS:

The Prototype detected a steady fluorescence increase in 5 sec in J774.1 after phagocytosis, using GE particles as a negative control in the presence of CyD. Furthermore, detection was possible at 10(4) cells/test, a concentration where the flow cytometer had difficulty for detection.

CONCLUSION:

The Prototype enables measurement of the phagocytic activity within a short period of time, even with a small sample amount, thus establishing the basic conditions of measurements of phagocytosis.
Assuntos
Palavras-chave
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fatores Ativadores de Macrófagos / Macrófagos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Anticancer Res Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Japão
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Fatores Ativadores de Macrófagos / Macrófagos Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Anticancer Res Ano de publicação: 2016 Tipo de documento: Article País de afiliação: Japão
...