Proposal of a New Electrode and Modality of Electrical Stimulation for Diaphragmatic Pacemakers.
Guellner Ghedini, Rodrigo; de Oliveira Paludo, Artur; Mariano, Rodrigo; Franco Silva, Éverton; Lise Simoneti, Lucas Elias; Dalla Giacomassa Rocha Thomaz, Leonardo; Feijó Andrade, Cristiano.
Afiliação
  • Guellner Ghedini R; Department of Physiotherapy, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil; Airway and Lung Laboratory, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil.
  • de Oliveira Paludo A; Airway and Lung Laboratory, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil.
  • Mariano R; Airway and Lung Laboratory, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil.
  • Franco Silva É; Airway and Lung Laboratory, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil.
  • Lise Simoneti LE; Airway and Lung Laboratory, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil.
  • Dalla Giacomassa Rocha Thomaz L; Airway and Lung Laboratory, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil.
  • Feijó Andrade C; Department of Thoracic Surgery, Hospital de Clínicas de Porto Alegre, Porto Alegre, Rio Grande do Sul, Brazil. Electronic address: cristianofa@gmail.com.
Ann Thorac Surg ; 105(6): 1851-1854, 2018 06.
Article em En | MEDLINE | ID: mdl-29477566
ABSTRACT

PURPOSE:

This report describes the implantation of a new design of intramuscular electrode to the diaphragm and introduces a new modality of electroventilation for diaphragmatic pacemaker. DESCRIPTION We used 22 Large White female pigs to test 68 electrodes that were implanted in the abdominal portion of the diaphragm by videolaparoscopy. Animals were submitted to a new protocol of electroventilation with an unpolarized current and a single-channel mode with 1 electrode for each hemidiaphragm during 3 hours. EVALUATION The electrodes were easily fixed to the diaphragm with the help of a deployment device designed for this purpose. There was no evidence of clinical complications such as pneumothorax or bleeding. No significant differences were observed in arterial blood gas analysis and serum lactate levels.

CONCLUSIONS:

The designed electrodes provided an excellent performance in connection between the electrical stimulator and the muscle tissue. The proposal for using an unpolarized current proved to be effective in stimulating both hemidiaphragms simultaneously with only a single channel of the electrical stimulator.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Diafragma / Estimulação Elétrica / Eletrodos Implantados Tipo de estudo: Diagnostic_studies / Guideline Limite: Animals Idioma: En Revista: Ann Thorac Surg Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Diafragma / Estimulação Elétrica / Eletrodos Implantados Tipo de estudo: Diagnostic_studies / Guideline Limite: Animals Idioma: En Revista: Ann Thorac Surg Ano de publicação: 2018 Tipo de documento: Article País de afiliação: Brasil
...