Comparison of the Panther Fusion and a laboratory-developed test targeting the envelope gene for detection of SARS-CoV-2.
Hogan, Catherine A; Sahoo, Malaya K; Huang, ChunHong; Garamani, Natasha; Stevens, Bryan; Zehnder, James; Pinsky, Benjamin A.
Afiliação
  • Hogan CA; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA; Clinical Virology Laboratory, Stanford Health Care, Stanford, CA, USA.
  • Sahoo MK; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA.
  • Huang C; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA.
  • Garamani N; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA.
  • Stevens B; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA; Clinical Virology Laboratory, Stanford Health Care, Stanford, CA, USA.
  • Zehnder J; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA.
  • Pinsky BA; Department of Pathology, Stanford University School of Medicine, Stanford, CA, USA; Clinical Virology Laboratory, Stanford Health Care, Stanford, CA, USA; Division of Infectious Diseases and Geographic Medicine, Department of Medicine, Stanford University School of Medicine, Stanford, CA, USA. Elect
J Clin Virol ; 127: 104383, 2020 06.
Article em En | MEDLINE | ID: mdl-32353760
BACKGROUND: Numerous nucleic acid amplification assays have recently received emergency use authorization (EUA) for the diagnosis of SARS-CoV-2 infection, and there is a need to assess their test performance relative to one another. OBJECTIVES: The aim of this study was to compare the test performance of the Hologic Panther Fusion SARS-CoV-2 assay targeting two regions of open reading frame 1ab (ORF1ab) to a high complexity molecular-based, laboratory-developed EUA from Stanford Health Care (SHC) targeting the SARS-CoV-2 envelope (E) gene. STUDY DESIGN: We performed a diagnostic comparison study by testing nasopharyngeal samples on the two assays. Assay agreement was assessed by overall percent agreement and Cohen's kappa coefficient. RESULTS: A total of 184 nasopharyngeal samples were tested using the two assays, of which 180 showed valid results and were included for the comparative analysis. Overall percent agreement between the assays was 98.3 % (95 % confidence interval (CI) 95.2-99.7) and kappa coefficient was 0.97 (95 % CI 0.93-1.0). One sample was detected on the SHC laboratory developed test (LDT) and not on the Panther Fusion, and had a Ct of 35.9. Conversely, 2 samples were detected on the Panther Fusion and not on the LDT, and had Ct values of 37.2 and 36.6. CONCLUSION: The Panther Fusion SARS-CoV-2 assay and the SHC LDT perform similarly on clinical nasopharyngeal swab specimens. Other considerations, including reagent availability, turnaround time, labor requirements, cost and instrument throughput should guide the decision of which assay to perform.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pneumonia Viral / Kit de Reagentes para Diagnóstico / Proteínas do Envelope Viral / Infecções por Coronavirus / Técnicas de Diagnóstico Molecular Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Virol Assunto da revista: VIROLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Pneumonia Viral / Kit de Reagentes para Diagnóstico / Proteínas do Envelope Viral / Infecções por Coronavirus / Técnicas de Diagnóstico Molecular Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Virol Assunto da revista: VIROLOGIA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Estados Unidos
...