Stability of dosomics features extraction on grid resolution and algorithm for radiotherapy dose calculation.
Placidi, L; Lenkowicz, J; Cusumano, D; Boldrini, L; Dinapoli, N; Valentini, V.
Afiliação
  • Placidi L; Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Roma, Italy; Istituto di Radiologia, Università Cattolica del Sacro Cuore, Roma, Italy.
  • Lenkowicz J; Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Roma, Italy. Electronic address: jacopo.lenkowicz@gmail.com.
  • Cusumano D; Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Roma, Italy; Istituto di Radiologia, Università Cattolica del Sacro Cuore, Roma, Italy.
  • Boldrini L; Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Roma, Italy; Istituto di Radiologia, Università Cattolica del Sacro Cuore, Roma, Italy.
  • Dinapoli N; Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Roma, Italy.
  • Valentini V; Fondazione Policlinico Universitario "A. Gemelli" IRCCS, Roma, Italy; Istituto di Radiologia, Università Cattolica del Sacro Cuore, Roma, Italy.
Phys Med ; 77: 30-35, 2020 Sep.
Article em En | MEDLINE | ID: mdl-32768918
ABSTRACT

PURPOSE:

Dosomics is a novel texture analysis method to parameterize regions of interest and to produce dose features that encode the spatial and statistical distribution of radiotherapy dose at higher resolution than organ-level dose-volume histograms. This study investigates the stability of dosomics features extraction, as their variation due to changes of grid resolution and algorithm dose calculation. MATERIAL AND

METHODS:

Dataset has been generated considering all the possible combinations of four grid resolutions and two algorithms dose calculation of 18 clinical delivered dose distributions, leading to a 144 3D dose distributions dataset. Dosomics features extraction has been performed with an in-house developed software. A total number of 214 dosomics features has been extracted from four different region of interest PTV, the two closest OARs and a RING structure. Reproducibility and stability of each extracted dosomic feature (Rfe, Sfe), have been analyzed in terms of intraclass correlation coefficient (ICC) and coefficient of variation.

RESULTS:

Dosomics features extraction was found reproducible (ICC > 0.99). Dosomic features, across the combination of grid resolutions and algorithms dose calculation, are more stable in the RING for all the considered feature's families. Sfe is higher in OARs, in particular for GLSZM features' families. Highest Sfe have been found in the PTV, in particular in the GLCM features' family.

CONCLUSION:

Stability and reproducibility of dosomics features have been evaluated for a representative clinical dose distribution case mix. These results suggest that, in terms of stability, dosomic studies should always perform a reporting of grid resolution and algorithm dose calculation.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radioterapia de Intensidade Modulada Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Radioterapia de Intensidade Modulada Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Phys Med Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2020 Tipo de documento: Article País de afiliação: Itália
...