Your browser doesn't support javascript.
loading
Trueness of ten intraoral scanners in determining the positions of simulated implant scan bodies.
Kim, Ryan Jin Young; Benic, Goran I; Park, Ji-Man.
Afiliação
  • Kim RJY; Department of Dental Science and Dental Research Institute, School of Dentistry, Seoul National University, Seoul, 03080, South Korea.
  • Benic GI; Clinic of Fixed and Removable Prosthodontics and Dental Material Science, Center for Dental Medicine, University of Zurich, 8032, Zurich, Switzerland.
  • Park JM; Department of Prosthodontics, College of Dentistry, Yonsei University, Seoul, 03722, South Korea. jimarn@yuhs.ac.
Sci Rep ; 11(1): 2606, 2021 01 28.
Article em En | MEDLINE | ID: mdl-33510317
Few investigations have evaluated the 3-dimensional (3D) accuracy of digital implant scans. The aim of this study was to evaluate the performance of 10 intraoral scanners (IOSs) (CEREC Omnicam, CEREC Primescan, CS 3600, DWIO, i500, iTero Element, PlanScan, Trios 2, Trios 3, and True Definition) in obtaining the accurate positions of 6 cylinders simulating implant scan bodies. Digital scans of each IOS were compared with the reference dataset obtained by means of a coordinate measuring machine. Deviation from the actual positions of the 6 cylinders along the XYZ axes and the overall 3D deviation of the digital scan were calculated. The type of IOSs and position of simulated cylindrical scan bodies affected the magnitude and direction of deviations on trueness. The lowest amount of deviation was found at the cylinder next to the reference origin, while the highest deviation was evident at the contralateral side for all IOSs (p < 0.001). Among the tested IOSs, the CEREC Primescan and Trios 3 had the highest trueness followed by i500, Trios 2, and iTero Element, albeit not statistically significant (p > 0.05), and the DWIO and PlasScan had the lowest trueness in partially edentulous mandible digital implant scans (p < 0.001).
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantes Dentários / Fotografia Dentária / Imageamento Tridimensional / Dentição Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Sci Rep Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Coréia do Sul

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantes Dentários / Fotografia Dentária / Imageamento Tridimensional / Dentição Tipo de estudo: Evaluation_studies Limite: Humans Idioma: En Revista: Sci Rep Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Coréia do Sul
...