Your browser doesn't support javascript.
loading
Real-time US/cone-beam CT fusion imaging for percutaneous ablation of small renal tumours: a technical note.
Monfardini, Lorenzo; Gennaro, Nicolò; Orsi, Franco; Della Vigna, Paolo; Bonomo, Guido; Varano, Gianluca; Solbiati, Luigi; Mauri, Giovanni.
Afiliação
  • Monfardini L; Division of Interventional Radiology, Fondazione Poliambulanza, 25124, Brescia, Italy. lorenzo.monfardini@poliambulanza.it.
  • Gennaro N; Departament of Biomedical Sciences, Humanitas University, 20072, Pieve Emanuele, Italy.
  • Orsi F; Departament of Radiology, Humanitas Clinical and Research Center, 20089, Rozzano, Italy.
  • Della Vigna P; Division of Interventional Radiology, IEO, European Institute of Oncology IRCCS, 20141, Milan, Italy.
  • Bonomo G; Division of Interventional Radiology, IEO, European Institute of Oncology IRCCS, 20141, Milan, Italy.
  • Varano G; Division of Interventional Radiology, IEO, European Institute of Oncology IRCCS, 20141, Milan, Italy.
  • Solbiati L; Division of Interventional Radiology, IEO, European Institute of Oncology IRCCS, 20141, Milan, Italy.
  • Mauri G; Departament of Biomedical Sciences, Humanitas University, 20072, Pieve Emanuele, Italy.
Eur Radiol ; 31(10): 7523-7528, 2021 Oct.
Article em En | MEDLINE | ID: mdl-33871707
ABSTRACT

INTRODUCTION:

Fusion imaging is gaining attention as an imaging technique to assist minimally invasive tumour ablation. Ultrasound (US) and computed tomography (CT) are the most common imaging modalities to guide thermal ablation of renal tumours, yet cone-beam CT (CBCT) has recently been described to successfully assist percutaneous renal interventions. Our goal was to evaluate primary technical success and correct lesion targeting of US/CBCT fusion imaging to guide the ablation of kidney masses < 2 cm in a small group of patients. MATERIAL AND

METHODS:

Six renal lesions (maximum diameter 11-17 mm) were treated with RFA in 5 different patients using real-time US/CBCT. Fusion imaging was used to identify and monitor tumour ablation. Demographics, tumour characteristics and mean serum creatinine levels were recorded before and after the procedure. Primary technical success and correct lesion targeting represented the main endpoints of the study. Primary technique efficacy was confirmed at 1-month and 3-month contrast-enhanced CT follow-ups.

RESULTS:

In all cases, a confident US/CBCT synchronisation was reached and allowed for a correct targeting and a successful percutaneous ablation. Primary technique efficacy was 100%. No recurrence was observed at the follow-up that ranged from 8 to 26 months (mean 16 months).

CONCLUSIONS:

US/CBCT fusion proved to be a viable method to precisely guide safe and effective percutaneous thermal ablation in patients with small renal tumours, especially when hardly detectable on US. KEY POINTS • US/CBCT fusion imaging for renal ablation is safe and feasible. • US/CBCT fusion imaging allows for an improved targeting and complete ablation of small RCC with poor US-conspicuity.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Cateter / Neoplasias Renais Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Ablação por Cateter / Neoplasias Renais Tipo de estudo: Observational_studies / Prognostic_studies Limite: Humans Idioma: En Revista: Eur Radiol Assunto da revista: RADIOLOGIA Ano de publicação: 2021 Tipo de documento: Article País de afiliação: Itália
...