Electrospray ionization ion mobility spectrometer with new tristate ion gating for improved sensitivity for compounds with lower ion mobility.
Thoben, C; Raddatz, C-R; Lippmann, M; Salehimoghaddam, Z; Zimmermann, S.
Afiliação
  • Thoben C; Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Appelstr. 9A, 30167, Hannover, Germany. Electronic address: thoben@geml.uni-hannover.de.
  • Raddatz CR; Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Appelstr. 9A, 30167, Hannover, Germany.
  • Lippmann M; Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Appelstr. 9A, 30167, Hannover, Germany.
  • Salehimoghaddam Z; Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Appelstr. 9A, 30167, Hannover, Germany.
  • Zimmermann S; Leibniz University Hannover, Institute of Electrical Engineering and Measurement Technology, Appelstr. 9A, 30167, Hannover, Germany.
Talanta ; 233: 122579, 2021 Oct 01.
Article em En | MEDLINE | ID: mdl-34215071
ABSTRACT
An electrospray is a dispersed nebula of charged droplets produced under the influence of a strong electric field. The charged droplets subsequently result in ions in the gas phase. Therefore, electrospray is a commonly used method for transferring liquids to the gas phase while ionizing its constituents at the same time. In this work, we investigate the performance of an electrospray ionization ion mobility spectrometer by varying the electric field strength in the desolvation region. In particular, we investigate a new tristate ion shutter with increased sensitivity for ions with higher molecular mass and lower ion mobility that are usually suppressed by classical Bradbury-Nielsen or Tyndall-Powell ion shutters when using short gating times as required for high resolving power. The electric field in the tristate ion shutter affects the optimal ratio of the electric field strengths in the drift and desolvation region. Furthermore, the solvent flow rate needs to be considered when setting the field strengths in the desolvation region. However, a higher electric field strength in the desolvation region affects the field at the emitter tip. For this reason, a smaller ratio of the drift field strength and the desolvation field strength is beneficial, especially since higher solvent flow rates require higher fields to initiate an electrospray. In this work, we use tetraoctylammonium bromide as an instrument standard and the fungicide metalaxyl, the herbicide isoproturon and the antibiotic cefuroxime as model compounds.