The missing link: Unlocking the power of cardiac rhythm monitoring device based QT interval detection.
Chu, Antony F; Rajagopal, Gautham; Sarkar, Shantanu.
Afiliação
  • Chu AF; Warren Alpert School of Medicine, Brown University, Rhode Island Hospital, 593 Eddy St, APC Building, Division of Cardiology, Providence, Rhode Island, USA.
  • Rajagopal G; Medtronic Inc, 8200 Coral Sea St. Mounds View, Mounds View, Minnesota, USA.
  • Sarkar S; Medtronic Inc, 8200 Coral Sea St. Mounds View, Mounds View, Minnesota, USA.
Pacing Clin Electrophysiol ; 45(3): 401-409, 2022 Mar.
Article em En | MEDLINE | ID: mdl-34964507
ABSTRACT

BACKGROUND:

The QT interval is of high clinical value as QT prolongation can lead to Torsades de Pointes (TdP) and sudden cardiac death. Insertable cardiac monitors (ICMs) have the capability of detecting both absolute and relative changes in QT interval. In order to determine feasibility for long-term ICM based QT detection, we developed and validated an algorithm for continuous long-term QT monitoring in patients with ICM.

METHODS:

The QT detection algorithm, intended for use in ICMs, is designed to detect T-waves and determine the beat-to-beat QT and QTc intervals. The algorithm was developed and validated using real-world ICM data. The performance of the algorithm was evaluated by comparing the algorithm detected QT interval with the manually annotated QT interval using Pearson's correlation coefficient and Bland Altman plot.

RESULTS:

The QT detection algorithm was developed using 144 ICM ECG episodes from 46 patients and obtained a Pearson's coefficient of 0.89. The validation data set consisted of 136 ICM recorded ECG segments from 76 patients with unexplained syncope and 104 ICM recorded nightly ECG segments from 10 patients with diabetes and Long QT syndrome. The QT estimated by the algorithm was highly correlated with the truth data with a Pearson's coefficient of 0.93 (p < .001), with the mean difference between annotated and algorithm computed QT intervals of -7 ms.

CONCLUSIONS:

Long-term monitoring of QT intervals using ICM is feasible. Proof of concept development and validation of an ICM QT algorithm reveals a high degree of accuracy between algorithm and manually derived QT intervals.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Síndrome do QT Longo / Torsades de Pointes Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Síndrome do QT Longo / Torsades de Pointes Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Revista: Pacing Clin Electrophysiol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Estados Unidos
...