Investigating the lateral dose response functions of point detectors in proton beams.
Kretschmer, Jana; Brodbek, Leonie; Looe, Hui Khee; van der Graaf, Emiel; Jan van Goethem, Marc; Kiewiet, Harry; Olivari, Francesco; Meyer, Christoph; Poppe, Björn; Brandenburg, Sytze.
Afiliação
  • Kretschmer J; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl-von-Ossietzky University Oldenburg, Germany.
  • Brodbek L; Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, The Netherlands.
  • Looe HK; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl-von-Ossietzky University Oldenburg, Germany.
  • van der Graaf E; Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, The Netherlands.
  • Jan van Goethem M; EBG MedAustron GmbH, Marie Curie-Straße 5, A-2700, Wiener Neustadt, Austria.
  • Kiewiet H; University Clinic for Medical Radiation Physics, Medical Campus Pius Hospital, Carl-von-Ossietzky University Oldenburg, Germany.
  • Olivari F; Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, The Netherlands.
  • Meyer C; Particle Therapy Research Center, University Medical Center Groningen, University of Groningen, The Netherlands.
  • Poppe B; Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, The Netherlands.
  • Brandenburg S; Particle Therapy Research Center, University Medical Center Groningen, University of Groningen, The Netherlands.
Phys Med Biol ; 67(14)2022 07 04.
Article em En | MEDLINE | ID: mdl-35697024
ABSTRACT
Objective. Point detector measurements in proton fields are perturbed by the volume effect originating from geometrical volume-averaging within the extended detector's sensitive volume and density perturbations by non-water equivalent detector components. Detector specific lateral dose response functionsK(x) can be used to characterize the volume effect within the framework of a mathematical convolution model, whereK(x) is the convolution kernel transforming the true dose profileD(x) into the measured signal profile of a detectorM(x). The aim of this work is to investigateK(x) for detectors in proton beams.Approach. TheK(x) for five detectors were determined by iterative deconvolution of measurements ofD(x) andM(x) profiles at 2 cm water equivalent depth of a narrow 150 MeV proton beam. Monte Carlo simulations were carried out for two selected detectors to investigate a potential energy dependence, and to study the contribution of volume-averaging and density perturbation to the volume effect.Main results. The Monte Carlo simulated and experimentally determinedK(x) agree within 2.1% of the maximum value. Further simulations demonstrate that the main contribution to the volume effect is volume-averaging. The results indicate that an energy or depth dependence ofK(x) is almost negligible in proton beams. While the signal reduction from a Semiflex 3D ionization chamber in the center of a gaussian shaped field with 2 mm sigma is 32% for photons, it is 15% for protons. When measuring the field with a microDiamond the trend is less pronounced and reversed with a signal reduction for protons of 3.9% and photons of 1.9%.Significance. The determinedK(x) can be applied to characterize the influence of the volume effect on detectors measured signal profiles at all clinical proton energies and measurement depths. The functions can be used to derive the actual dose distribution from point detector measurements.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Contexto em Saúde: 1_ASSA2030 Problema de saúde: 1_financiamento_saude Assunto principal: Prótons / Radiometria Tipo de estudo: Health_economic_evaluation Idioma: En Revista: Phys Med Biol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Contexto em Saúde: 1_ASSA2030 Problema de saúde: 1_financiamento_saude Assunto principal: Prótons / Radiometria Tipo de estudo: Health_economic_evaluation Idioma: En Revista: Phys Med Biol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Alemanha
...