In Vivo Models for the Performance and Safety of BAT-90, a Novel 90-Yttrium-based Internal Radiotherapy Platform.
In Vivo
; 36(5): 2052-2060, 2022.
Article
em En
| MEDLINE
| ID: mdl-36099118
BACKGROUND/AIM: BAT-90 is an innovative active implantable device designed for the irradiation of unresectable tumors (e.g., liver cancer) or surgical tumor beds, based on the combination of Yttrium-90 beta-emitting microspheres and a tissue adhesive hydrogel, currently used in cardio-vascular surgery. The rationale behind BAT-90 is to localize the Yttrium-90 activity on the administration site, while minimizing its body dispersion. MATERIALS AND METHODS: The effective induction of necrosis in the target injection area was tested in a pig liver model, whereas the safety of BAT-90 was assessed and demonstrated in biocompatibility tests for acute systemic toxicity, intracutaneous reactivity, delayed hypersensitivity and subcutaneous implantation. RESULTS: BAT-90 administration induced necrosis into the target site, while the safety experiments in the treated animals highlighted results very similar to the controls. CONCLUSION: BAT-90 could be considered as a safe and innovative treatment option for inoperable solid tumors of the liver.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Radioisótopos de Ítrio
/
Neoplasias Hepáticas
Limite:
Animals
Idioma:
En
Revista:
In Vivo
Assunto da revista:
NEOPLASIAS
Ano de publicação:
2022
Tipo de documento:
Article