Agreement of somatic and renal near-infrared spectroscopy with reference blood samples during a controlled hypoxia sequence: a healthy volunteer study.
De Keijzer, Ilonka N; Massari, Dario; Niezen, Caren K; Bokkers, Reinoud P H; Vos, Jaap Jan; Scheeren, Thomas W L.
Afiliação
  • De Keijzer IN; Department of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands. i.n.de.keijzer@umcg.nl.
  • Massari D; Department of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Niezen CK; Department of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Bokkers RPH; Department of Radiology, Medical Imaging Center, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Vos JJ; Department of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
  • Scheeren TWL; Department of Anesthesiology, University Medical Center Groningen, University of Groningen, Groningen, The Netherlands.
J Clin Monit Comput ; 37(3): 805-814, 2023 06.
Article em En | MEDLINE | ID: mdl-36463540
ABSTRACT

PURPOSE:

O3® Regional Oximetry (Masimo Corporation, California, USA) is validated for cerebral oximetry. We aimed to assess agreement of somatic and renal near-infrared spectroscopy with reference blood samples.

METHODS:

O3 sensors were placed bilaterally on the quadriceps and flank of 26 healthy volunteers. A stepped, controlled hypoxia sequence was performed by adding a mixture of nitrogen and room air to the breathing circuit. O3-derived oxygen saturation values were obtained at baseline and at six decremental saturation levels (5% steps). Blood samples (radial artery, iliac vein (somatic reference) and renal vein) were obtained at each step. Reference values were calculated as 0.7 × venous saturation + 0.3 × arterial saturation. The agreement between O3-derived values with blood reference values was assessed by calculating root-mean-square error accuracy and Bland-Altman plots.

RESULTS:

The root-mean-square error accuracy was 6.0% between quadriceps oxygen saturation and somatic reference values. The mean bias was 0.8%, with limits of agreement from -7.7 to 9.3%. These were 5.1% and 0.6% (-8.3 to 9.5%) for flank oxygen saturation and somatic reference values, respectively, and 7.7% and -4.9% (-15.0 to 5.2%) for flank oxygen saturation and renal reference values. The kidney depth was 3.1 ± 0.9 cm below the skin.

CONCLUSION:

O3 regional oximetry can be used on the quadriceps and flank to monitor somatic saturation, yet has a saturation-level dependent bias. O3-derived values obtained at the flank underestimated renal reference values. Additionally, it is unlikely that the flank sensors did directly measure renal tissue. TRIAL REGISTRATION Clinicaltrials.gov (NCT04584788) registered October 6th, 2020.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oximetria / Espectroscopia de Luz Próxima ao Infravermelho Limite: Humans Idioma: En Revista: J Clin Monit Comput Assunto da revista: INFORMATICA MEDICA / MEDICINA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oximetria / Espectroscopia de Luz Próxima ao Infravermelho Limite: Humans Idioma: En Revista: J Clin Monit Comput Assunto da revista: INFORMATICA MEDICA / MEDICINA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Holanda
...