ABSTRACT

BACKGROUND:

Hyperkalaemia is managed in the emergency department (ED) following measurement of potassium results by blood gas analysers (BGA) or laboratory analysers (LAB).

AIMS:

To determine the prevalence of clinically significant differences between BGA and LAB potassium results and the impact on ED hyperkalaemia management.

METHODS:

Retrospective analysis of time-matched ED BGA and LAB potassium samples from 2019 to 2020 (taken within 15 min, one or both results ≥6.0 mmol/L). Mean differences and 95% limits of agreement (LoA) were determined for pairs with one or both results ≥6.0 mmol/L and a separate 500 consecutive sample pairs.

RESULTS:

Four hundred eighty-eight matched BGA and LAB samples met the inclusion criteria. Of these, 201 (41.2%) differed by ≤0.5 mmol/L, 169 (34.6%) included a haemolysed LAB sample, and 12 (2.5%) had an unreportable BGA sample. One hundred six (21.7%) pairs differed by >0.5 mmol/L, and 60/106 (57%) had normal LAB potassium results, but BGA indicated moderate/severe hyperkalaemia (two of these pairs received hyperkalaemia treatment). Of patients with a haemolysed LAB sample, or where pairs differed by >0.5 mmol, 48 were treated with insulin and five (10.4%) experienced hypoglycaemia. Mean differences and LoA for pairs with LAB results <6.0 mmol/L but BGA ≥6.0 mmol/L demonstrated unacceptable agreement, with 18 (25.7%) BGA results exceeding 8.0 mmol/L.

CONCLUSIONS:

Potentially significant discordance may occur between BGA and LAB potassium results. Clinicians need to be aware of factors impacting both analytical methods' accuracy (such as poor venepuncture or sample handling, (K) EDTA interference) and undetectable haemolysis with BGA measurements. We recommend BGA hyperkalaemia be confirmed with LAB results using a non-haemolysed sample where time permits.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Potássio / Hiperpotassemia Tipo de estudo: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Intern Med J Assunto da revista: MEDICINA INTERNA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Austrália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Potássio / Hiperpotassemia Tipo de estudo: Diagnostic_studies / Prognostic_studies / Risk_factors_studies Limite: Humans Idioma: En Revista: Intern Med J Assunto da revista: MEDICINA INTERNA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Austrália
...