ABSTRACT

BACKGROUND:

Platelet function testing in cats allows determination of clopidogrel effect. Plateletworks assesses aggregation based on decreasing platelet counts on hematology analyzers in response to agonists. It has not been validated for the IDEXX ProCyte Dx analyzer. Ideal time to perform analysis and the utility of other platelet parameters have not been fully assessed.

OBJECTIVES:

To validate Plateletworks ADP on the ProCyte Dx, to investigate the utility of various platelet parameters using Plateletworks ADP, and determine the ideal time to perform analysis. ANIMALS Twenty healthy cats recruited from the general population used for transference of reference intervals to a new analyzer, and 10 cats receiving clopidogrel to determine clopidogrel effect.

METHODS:

Plateletworks ADP using the ProCyte Dx and ADVIA 2120i analyzer was run simultaneously in both healthy cats and cats receiving clopidogrel, and CBC results at different timepoints were compared between analyzers.

RESULTS:

Aggregation was significantly different (P < .001) between analyzers. Cohen's kappa showed almost perfect agreement for determination of clopidogrel effect, and the area under the curve of the receiver operating characteristic was 1.0. Lower limits of the aggregation reference interval in healthy cats were 28.8% on the ProCyte Dx and 12.5% on the ADVIA 2120i. Coefficients of variation for platelet parameters were not different between analyzers. No significant changes in mean platelet volume, plateletcrit, large platelets, and mean platelet component were identified. No significant change in aggregation was observed within the first hour after phlebotomy. CONCLUSIONS AND CLINICAL IMPORTANCE Our study validated the Plateletworks ADP system on the ProCyte Dx analyzer. Samples may be analyzed up to 1 h after collection.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Função Plaquetária / Agregação Plaquetária Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: J Vet Intern Med Assunto da revista: MEDICINA INTERNA / MEDICINA VETERINARIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Canadá

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Testes de Função Plaquetária / Agregação Plaquetária Tipo de estudo: Prognostic_studies Limite: Animals Idioma: En Revista: J Vet Intern Med Assunto da revista: MEDICINA INTERNA / MEDICINA VETERINARIA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Canadá
...