Your browser doesn't support javascript.
loading
Simultaneous determination of colistin sulfate and tigecycline in human plasma by liquid chromatography-tandem mass spectrometry method.
Ma, Ying-Chao; Wu, Xi-Kun; Yang, Xiu-Ling; Zhang, Zhi-Qing.
Afiliação
  • Ma YC; Department of Pharmacy, Second Hospital of Hebei Medical University, No. 215 of Heping West Road,Xinhua District, Shijiazhuang, 050000, China.
  • Wu XK; Department of Pharmacy, Second Hospital of Hebei Medical University, No. 215 of Heping West Road,Xinhua District, Shijiazhuang, 050000, China.
  • Yang XL; Department of Pharmacy, Second Hospital of Hebei Medical University, No. 215 of Heping West Road,Xinhua District, Shijiazhuang, 050000, China. yangxuilingyxl0d@126.com.
  • Zhang ZQ; Department of Pharmacy, Second Hospital of Hebei Medical University, No. 215 of Heping West Road,Xinhua District, Shijiazhuang, 050000, China. zhangzhiqingzzqo82@126.com.
BMC Chem ; 18(1): 12, 2024 Jan 13.
Article em En | MEDLINE | ID: mdl-38218863
ABSTRACT

OBJECTIVE:

To establish a high-performance liquid chromatography-tandem mass spectrometry method (HPLC-MS/MS) to simultaneously determine colistin sulfate and tigecycline in human plasma.

METHODS:

Polymyxin B1 internal standard (20 µL) was added into 200 µL of plasma sample. The samples were treated with methanol-5% trichloroacetic acid (5050, V/V) solution, and the protein precipitation method was adopted for post-injection analysis. The chromatographic column was a Dikma C18 (4.6 mm × 150 mm, 5 µm). For the mobile phase, 0.1% formic acid in aqueous solution was used for phase A, 0.1% formic acid in acetonitrile solution for phase B, and gradient elution was also applied. The flow rate was 0.8 mL/min, the column temperature was 40 °C, and the injection volume was 10 µL; Electrospray ionization and multiple reaction ion monitoring were adopted and scanned by the HPLC-MS/MS positive ion mode.

RESULTS:

The endogenous impurities in the plasma had no interference in the determination of the analytes. There existed a good linear relationship of colistin sulfate within the range of 0.1-10 µg/mL (R2 = 0.9986), with the lower limit of quantification (LLOQ) of 0.1 µg/mL. There existed a good linear relationship of tigecycline within the range of 0.05-5 µg/ mL (R2 = 0.9987), with the LLOQ of 0.05 µg/mL. The intra- and inter-day relative standard deviations of colistin sulfate and tigecycline were both less than 15%, and the accuracy was between 88.21% and 108.24%. The extraction had good stability, the extraction recovery rate was 87.75-91.22%, and the matrix effect was 99.40-105.26%.

CONCLUSION:

This study successfully established a method for simultaneously detecting colistin sulfate and tigecycline plasma concentrations. The method was simple, rapid, and highly sensitive and could be applied for therapeutic medication monitoring.
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: BMC Chem Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: BMC Chem Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China
...