First-in-human validation of a DROP-IN ß-probe for robotic radioguided surgery: defining optimal signal-to-background discrimination algorithm.
Eur J Nucl Med Mol Imaging
; 51(10): 3098-3108, 2024 Aug.
Article
em En
| MEDLINE
| ID: mdl-38376805
ABSTRACT
PURPOSE:
In radioguided surgery (RGS), radiopharmaceuticals are used to generate preoperative roadmaps (e.g., PET/CT) and to facilitate intraoperative tracing of tracer avid lesions. Within RGS, there is a push toward the use of receptor-targeted radiopharmaceuticals, a trend that also has to align with the surgical move toward minimal invasive robotic surgery. Building on our initial ex vivo evaluation, this study investigates the clinical translation of a DROP-IN ß probe in robotic PSMA-guided prostate cancer surgery.METHODS:
A clinical-grade DROP-IN ß probe was developed to support the detection of PET radioisotopes (e.g., 68 Ga). The prototype was evaluated in 7 primary prostate cancer patients, having at least 1 lymph node metastases visible on PSMA-PET. Patients were scheduled for radical prostatectomy combined with extended pelvic lymph node dissection. At the beginning of surgery, patients were injected with 1.1 MBq/kg of [68Ga]Ga-PSMA. The ß probe was used to trace PSMA-expressing lymph nodes in vivo. To support intraoperative decision-making, a statistical software algorithm was defined and optimized on this dataset to help the surgeon discriminate between probe signals coming from tumors and healthy tissue.RESULTS:
The DROP-IN ß probe helped provide the surgeon with autonomous and highly maneuverable tracer detection. A total of 66 samples (i.e., lymph node specimens) were analyzed in vivo, of which 31 (47%) were found to be malignant. After optimization of the signal cutoff algorithm, we found a probe detection rate of 78% of the PSMA-PET-positive samples, a sensitivity of 76%, and a specificity of 93%, as compared to pathologic evaluation.CONCLUSION:
This study shows the first-in-human use of a DROP-IN ß probe, supporting the integration of ß radio guidance and robotic surgery. The achieved competitive sensitivity and specificity help open the world of robotic RGS to a whole new range of radiopharmaceuticals.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Neoplasias da Próstata
/
Algoritmos
/
Cirurgia Assistida por Computador
/
Procedimentos Cirúrgicos Robóticos
Limite:
Aged
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Eur J Nucl Med Mol Imaging
Assunto da revista:
MEDICINA NUCLEAR
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Itália