Very-early detection of atrial fibrillation after ablation evaluated by a wearable ECG-patch predicts late blanking period recurrence: Preliminary data from a prospective registry.
Marques Antunes, Miguel; Silva Cunha, Pedro; Lacerda Teixeira, Bárbara; Portugal, Guilherme; Valente, Bruno; Lousinha, Ana; Delgado, Ana Sofia; Alves, Sandra; Guerra, Cátia; Cruz Ferreira, Rui; Martins Oliveira, Mário.
Afiliação
  • Marques Antunes M; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Silva Cunha P; Clinical Pharmacology and Therapeutics Unit - Centro Cardiovascular Da Universidade de Lisboa (CCUL@RISE), Faculdade de Medicina Da Universidade de Lisboa.
  • Lacerda Teixeira B; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Portugal G; Faculdade de Medicina, CCUL, Universidade de Lisboa, Lisbon, Portugal.
  • Valente B; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Lousinha A; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Delgado AS; Faculdade de Medicina, CCUL, Universidade de Lisboa, Lisbon, Portugal.
  • Alves S; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Guerra C; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Cruz Ferreira R; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
  • Martins Oliveira M; Cardiology Department, Hospital Santa Marta, Centro Hospitalar Universitário de Lisboa Central, Centro Clínico Académico de Lisboa (CCAL), Lisbon, Portugal.
Int J Cardiol Heart Vasc ; 51: 101369, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38420510
ABSTRACT

Introduction:

Atrial fibrillation (AF) ablation represents a safe and effective procedure to restore sinus rhythm. The idea that post-procedural AF episodes - during the blanking period - are not considered treatment failure has been increasingly challenged. The E-Patch, a single-use adhesive electrode, facilitates extended continuous ECG monitoring for 120 h. This pilot study aims to assess the effectiveness of this ambulatory monitoring device and investigate whether very-early AF recurrence correlates with delayed blanking period ablation outcomes.

Methods:

We conducted a single-center, prospective, longitudinal study, including consecutive post-ablation patients monitored with the E-patch. The ability of the device to continuously record was analyzed, as well as the occurrence of AF episodes during external 7-day loop-recorder in the 2nd-month post-ablation.

Results:

We included 40 patients, median age 62 years (IQR 56-70). E-Patch monitoring was obtained for a median of 118 h (IQR 112-120), with no discomfort nor interpretation artefacts. Very-early AF recurrence was detected in 11 (27.5 %) patients, with a median AF burden of 7 % (IQR 6 %-33 %). Late-blanking period AF was detected in 13 (33 %) of the external 7-day loop recordings. Of the 11 patients that had very-early AF recurrence, 10 (91 %) had late-blanking AF. Very-early AF detection showed 77 % (95 % CI 64 %-90 %) sensitivity and 96 % (95 % CI 90-100 %) specificity in predicting late-blanking AF, with a non-parametric ROC curve AUC of 0.903 (95 % 0.797--1.0).

Conclusion:

The E-Patch was able to detect very-early AF during an extended period. Very-early AF detection emerges as a predictor of AF recurrence during the late blanking period post-ablation.