Iso-E-Codelock: A Rebuilding-free Electrochemical Chip with a Customizable Decoding Probe for Real-Time and Portable Pathogen Diagnostics.
Angew Chem Int Ed Engl
; 63(19): e202400340, 2024 05 06.
Article
em En
| MEDLINE
| ID: mdl-38497899
ABSTRACT
In order to realize portable pathogen diagnostics with easier quantitation, digitization and integration, we develop a ready-to-use electrochemical sensing strategy (Iso-E-Codelock) for real-time detection of isothermal nucleic acid amplification. Bridged by a branched DNA as codelock, the isothermal amplicon is transduced into increased current of an electrochemical probe, holding multiple advantages of high sensitivity, high selectivity, signal-on response, "zero" background and one-pot operation. Through a self-designed portable instrument (BioAlex PHE-T), the detection can be implemented on a multichannel microchip and output real-time amplification curves just like an expensive commercial PCR machine. The microchip is a rebuilding-free and disposable component. The branch codelock probe can be customized for different targets and designs. Such high performance and flexibility have been demonstrated utilizing four virus (SARS-CoV-2, African swine fever, FluA and FluB) genes as targets, and two branch (3-way and 4-way) DNAs as codelock probes.