A fluid-structure interaction simulation on the impact of transcatheter micro ventricular assist devices on aortic valves.
Wang, Yitao; He, Feng; Hao, Pengfei; Zhang, Xiwen.
Afiliação
  • Wang Y; Department of Engineering Mechanics, School of Aerospace Engineering, Tsinghua University, Beijing 100084, China.
  • He F; Department of Engineering Mechanics, School of Aerospace Engineering, Tsinghua University, Beijing 100084, China.
  • Hao P; Department of Engineering Mechanics, School of Aerospace Engineering, Tsinghua University, Beijing 100084, China; Applied Mechanics Laboratory, Department of Engineering Mechanics, Tsinghua University, Beijing 100084, PR China.
  • Zhang X; Department of Engineering Mechanics, School of Aerospace Engineering, Tsinghua University, Beijing 100084, China. Electronic address: zhangxiw@tsinghua.edu.cn.
Comput Methods Programs Biomed ; 254: 108270, 2024 Sep.
Article em En | MEDLINE | ID: mdl-38878358
ABSTRACT
BACKGROUND AND

OBJECTIVE:

The implantation of ventricular assist devices (VADs) has become an important treatment option for patients with heart failure. Aortic valve insufficiency is a common complication caused by VADs implantation. Currently, there is very little quantitative research on the effects of transcatheter micro VADs or the intervention pumps on the aortic valves.

METHODS:

In this study, the multi-component arbitrary Lagrange-Eulerian method is used to perform fluid-structure interaction simulations of the aortic valve model with and without intervention pumps. The effects of intervention pumps implantation on the opening area of the aortic valves, the stress distribution, and the flow characteristics are quantitatively analyzed. Statistical results are consistent with clinical guidelines and experiments.

RESULTS:

The implantation of intervention pumps leads to the valve insufficiency and causes weak valve regurgitation. In the short-term treatment, the valve regurgitation is within a controllable range. The distribution and variation of stress on the leaflets are also affected by intervention pumps. The whirling flow in the flow direction affects the closing speed of the aortic valves and optimizes the stress distribution of the valves. In the models with whirling flow, the effects of intervention pumps implantation on valve motion and stress distribution differ from those without whirling flow. However, the valve insufficiency and valve regurgitation caused by intervention pumps still exist in the models with whirling flow. Conventional artificial bioprosthetic valves have limited effectiveness in treating the valve diseases caused by intervention pumps implantation.

CONCLUSIONS:

This study quantitatively investigates the impact of intervention pumps on the aortic valves, and investigates the effect of blood rotation on the valve behavior, which is a gap in previous research. We suggest that in the short-term treatment, the implantation of intervention pumps has limited impact on aortic valves, caution should be exercised against valve regurgitation issues caused by intervention pumps.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Simulação por Computador / Coração Auxiliar / Modelos Cardiovasculares Limite: Humans Idioma: En Revista: Comput Methods Programs Biomed Assunto da revista: INFORMATICA MEDICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Valva Aórtica / Simulação por Computador / Coração Auxiliar / Modelos Cardiovasculares Limite: Humans Idioma: En Revista: Comput Methods Programs Biomed Assunto da revista: INFORMATICA MEDICA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China
...