Intervisit Reproducibility of Foveal Cone Density Metrics.
Adhan, Iniya; Warr, Emma; Grieshop, Jenna; Kreis, Joseph; Nikezic, Danica; Walesa, Ashleigh; Hemsworth, Katherine; Cooper, Robert F; Carroll, Joseph.
Afiliação
  • Adhan I; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Warr E; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Grieshop J; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Kreis J; Joint Department of Biomedical Engineering, Marquette University and Medical College of Wisconsin, Milwaukee, WI, USA.
  • Nikezic D; Department of Cell Biology, Neurobiology & Anatomy, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Walesa A; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Hemsworth K; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Cooper RF; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
  • Carroll J; Department of Ophthalmology & Visual Sciences, Medical College of Wisconsin, Milwaukee, WI, USA.
Transl Vis Sci Technol ; 13(6): 18, 2024 Jun 03.
Article em En | MEDLINE | ID: mdl-38913007
ABSTRACT

Purpose:

To assess longitudinal reproducibility of metrics of foveal density (peak cone density [PCD], cone density centroid [CDC], and 80th percentile centroid area) in participants with normal vision.

Methods:

Participants (n = 19; five male and 14 female) were imaged at two time points (average interval of 3.2 years) using an adaptive optics scanning light ophthalmoscope (AOSLO). Foveally centered regions of interest (ROIs) were extracted from AOSLO montages. Cone coordinate matrices were semiautomatically derived for each ROI, and cone mosaic metrics were calculated.

Results:

On average, there were no significant changes in cone mosaic metrics between visits. The average ± SD PCD was 187,000 ± 20,000 cones/mm2 and 189,000 ± 21,700 cones/mm2 for visits 1 and 2, respectively (P = 0.52). The average ± SD density at the CDC was 183,000 ± 19,000 cones/mm2 and 184,000 ± 20,800 cones/mm2 for visits 1 and 2, respectively (P = 0.78). The average ± SD 80th percentile isodensity contour area was 15,400 ± 1800 µm2 and 15,600 ± 1910 µm2 for visits 1 and 2, respectively (P = 0.57).

Conclusions:

Foveal cone mosaic density metrics were highly reproducible in the cohort examined here, although further study is required in more diverse populations. Translational Relevance Determination of the normative longitudinal changes in foveal cone topography is key for evaluating longitudinal measures of foveal cone topography in patients with progressive retinal dystrophies.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Células Fotorreceptoras Retinianas Cones / Fóvea Central Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Transl Vis Sci Technol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Células Fotorreceptoras Retinianas Cones / Fóvea Central Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Transl Vis Sci Technol Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos
...