Critical Hemorrhage Caused by a Size-Mismatched Extracorporeal Membrane Oxygenation Cannula in a Patient with Myotonic Dystrophy Type 1: A Case Report and Literature Review.
Shin, Changsik; Yoo, Kwon Cheol; Kim, Dae Hoon.
Afiliação
  • Shin C; Department of Surgery, Uijeongbu Eulji Medical Center, Eulji University, Uijeongbu 11759, Republic of Korea.
  • Yoo KC; Department of Surgery, College of Medicine, Chungbuk National University, Cheongju 28644, Republic of Korea.
  • Kim DH; Department of Surgery, Chungbuk National University Hospital, Cheongju 28644, Republic of Korea.
Medicina (Kaunas) ; 60(6)2024 Jun 12.
Article em En | MEDLINE | ID: mdl-38929586
ABSTRACT
Background and

Objective:

Although extracorporeal membrane oxygenation (ECMO) is an essential life-saving technique for patients with refractory cardiopulmonary shock, it can be fatal in certain cases. Case Presentation A 19-year-old girl treated with ECMO presented with acute limb ischemia 2 days after cannula removal. The decannulation was performed percutaneously by an interventional cardiologist, and the vascular surgery department was consulted after the patient developed symptoms. The first suspected diagnosis was thrombosis due to incorrect use of the closure device. However, the artery had ruptured due to the insertion of a catheter with a cannula that was larger than the patient's artery. Management and

Outcome:

Fortunately, excessive bleeding due to the size-mismatched cannula was prevented by an unintentional complication of the closing device, which saved the patient's life. She underwent a right common femoral artery thrombectomy and patch angioplasty. Hospital guidelines have changed regarding the surgical removal of ECMO cannulas.

Discussion:

This report aims to highlight the importance of two aspects that are critical to a successful

outcome:

individualized cannula selection followed by precise insertion and removal and postoperative evaluation of a patient's final status.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigenação por Membrana Extracorpórea / Cânula / Hemorragia / Distrofia Miotônica Limite: Adult / Female / Humans Idioma: En Revista: Medicina (Kaunas) / Medicina (Kaunas. Online) Assunto da revista: MEDICINA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigenação por Membrana Extracorpórea / Cânula / Hemorragia / Distrofia Miotônica Limite: Adult / Female / Humans Idioma: En Revista: Medicina (Kaunas) / Medicina (Kaunas. Online) Assunto da revista: MEDICINA Ano de publicação: 2024 Tipo de documento: Article
...