Your browser doesn't support javascript.
loading
Low-contrast lesion detection in neck CT: a multireader study comparing deep learning, iterative, and filtered back projection reconstructions using realistic phantoms.
Bellmann, Quirin; Peng, Yang; Genske, Ulrich; Yan, Li; Wagner, Moritz; Jahnke, Paul.
Afiliação
  • Bellmann Q; Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Charitéplatz 1, 10117, Berlin, Germany.
  • Peng Y; Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Charitéplatz 1, 10117, Berlin, Germany.
  • Genske U; Department of Radiology, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, 1095 Jiefang Avenue, Wuhan, 430030, Hubei Province, China.
  • Yan L; Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Charitéplatz 1, 10117, Berlin, Germany.
  • Wagner M; Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Charitéplatz 1, 10117, Berlin, Germany.
  • Jahnke P; Department of Radiology, Charité-Universitätsmedizin Berlin, Corporate Member of Freie Universität Berlin, Humboldt-Universität zu Berlin, and Berlin Institute of Health, Charitéplatz 1, 10117, Berlin, Germany.
Eur Radiol Exp ; 8(1): 84, 2024 Jul 24.
Article em En | MEDLINE | ID: mdl-39046565
ABSTRACT

BACKGROUND:

Computed tomography (CT) reconstruction algorithms can improve image quality, especially deep learning reconstruction (DLR). We compared DLR, iterative reconstruction (IR), and filtered back projection (FBP) for lesion detection in neck CT.

METHODS:

Nine patient-mimicking neck phantoms were examined with a 320-slice scanner at six doses 0.5, 1, 1.6, 2.1, 3.1, and 5.2 mGy. Each of eight phantoms contained one circular lesion (diameter 1 cm; contrast -30 HU to the background) in the parapharyngeal space; one phantom had no lesions. Reconstruction was made using FBP, IR, and DLR. Thirteen readers were tasked with identifying and localizing lesions in 32 images with a lesion and 20 without lesions for each dose and reconstruction algorithm. Receiver operating characteristic (ROC) and localization ROC (LROC) analysis were performed.

RESULTS:

DLR improved lesion detection with ROC area under the curve (AUC) 0.724 ± 0.023 (mean ± standard error of the mean) using DLR versus 0.696 ± 0.021 using IR (p = 0.037) and 0.671 ± 0.023 using FBP (p < 0.001). Likewise, DLR improved lesion localization, with LROC AUC 0.407 ± 0.039 versus 0.338 ± 0.041 using IR (p = 0.002) and 0.313 ± 0.044 using FBP (p < 0.001). Dose reduction to 0.5 mGy compromised lesion detection in FBP-reconstructed images compared to doses ≥ 2.1 mGy (p ≤ 0.024), while no effect was observed with DLR or IR (p ≥ 0.058).

CONCLUSION:

DLR improved the detectability of lesions in neck CT imaging. Dose reduction to 0.5 mGy maintained lesion detectability when denoising reconstruction was used. RELEVANCE STATEMENT Deep learning enhances lesion detection in neck CT imaging compared to iterative reconstruction and filtered back projection, offering improved diagnostic performance and potential for x-ray dose reduction. KEY POINTS Low-contrast lesion detectability was assessed in anatomically realistic neck CT phantoms. Deep learning reconstruction (DLR) outperformed filtered back projection and iterative reconstruction. Dose has little impact on lesion detectability against anatomical background structures.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Imagens de Fantasmas / Aprendizado Profundo / Neoplasias de Cabeça e Pescoço Limite: Humans Idioma: En Revista: Eur Radiol Exp Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Alemanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Imagens de Fantasmas / Aprendizado Profundo / Neoplasias de Cabeça e Pescoço Limite: Humans Idioma: En Revista: Eur Radiol Exp Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Alemanha
...